DIGITAL MULTIMETERS 7561 & 7562 Digital Multimeters RECORDERS
RECORDERS
WAVEFORM
MEASURING
INSTRUMENTS
DIGITAL
MEASURING
INSTRUMENTS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
7561 & 7562
Digital Multimeters
7561 (6-1/2 digits)
213 × 88 × 330 mm 3 kg
(8-3/8 × 3-1/2 × 13" 6.6 lbs)
The 7560 (6-1/2 digit display) Series is a high-accuracy,
high-speed sampling bench-top digital multimeter lineup using a YOKOGAWA-original A/D converter (feedback pulse
width modulation method), which features excellent stability, high noise immunity and linearity. In addition to the basic features, an optional IC memory card capable of storing
8,000 data, GP-IB or RS-232-C interface and versatile measurement functions allow these multimeters to be used in diverse fields of applications, from bench to system uses.
7562 (6-1/2 digits)
213 × 88 × 330 mm 3 kg
(8-3/8 × 3-1/2 × 13" 6.6 lbs)
■ High-speed sampling
• 7560 Series: 333 times/s
The feedback PWM A/D converter, which includes a
new counting system, allows a high-speed response
without changing any excellent features.
The sampling rate of 40 times/s can be obtained even
with a 5-1/2 digit display. The 7560 Series offer high system throughput with a large memory capacity, highspeed auto-ranging, programmable sampling interval,
and integrating time.
■ GP-IB or RS-232-C interface (standard)
■ High accuracy, and excellent stability and linearity
• 7560 Series
DC voltage accuracy : ±0.003% of rdg ± 15 digits
(90 days, 23 ± 5°C)
OHM accuracy
: ±0.009% of rdg ± 30 digits
(90 days, 23 ± 5°C)
■ A large memory capacity of 8,000 data by using IC
memory card
The 7560 Series can store 1,000 measured data by using
internal memory (or 8,000 data by using an optional IC
memory card).
This allows high-speed sampling without being affected
by transmission capacity via a communication bus.
Data can then be transferred to a controller after completing measurements.
These storage functions allow DMM’s to be used as
high-speed data logger.
The following functions are available with communication interface.
● Functions executed by panel key operation
● Measured data output
● Panel-set data output
● Stored data output
● Status byte output
■ An input-site change between front and rear panels
7560 Series has the input terminal for the voltage and resistance between front and rear panels.
15
DIGITAL MULTIMETERS
FEATURES
RECORDERS
WAVEFORM
MEASURING
INSTRUMENTS
DIGITAL
MEASURING
INSTRUMENTS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
■ MEASURING RANGE
• Auto Range Mode
When the display data increases more than 20,000/200,000,
the range is automatically changed (increased), whereas
when the display data goes down to less than 1,800/18,000,
the range is automatically changed (decreased).
■ TIME DELAY FUNCTIONS (TD)
The time interval between a trigger signal generation and the
sampling can be set from 0 to 60 minutes.
This allows accurate measurements when a signal source
has dead-time or the measurement circuit shows a first-order
lag response.
Input
• Manual Range Mode
If the measuring range exceeds upper-limit setpoint, –oL–
(overload) is displayed.
Response of
measuring
circuit
Time delay
■ SAMPLING FUNCTIONS
• N Reading Mode (N RDGS)
Data are sampled in the cycles set by the integral time and
measuring intervals.
■ TRIGGER FUNCTIONS
The following three ways are available for generating trigger
signals.
• Press the TRIG key on the front panel.
• Input a contact signal or TTL logic signal to the I/O signal
connector (pin No. 1).
• Send the “E” or GET command through GP-IB or RS-232-C.
In STORE mode, triggering performs three types of operations according to measurement modes as follows:
• In AUTO mode, data are written leaving NS-1 data before
generating trigger signals and the STORE mode is automatically turned OFF when memory capacity is full. Amounts of
data that are written depend on the built-in memory or IC
card memory. (PRE-TRIGGER FUNCTION)
• In SINGLE mode, sampled data is stored every time a trigger signal is generated. If NS data that are preset are stored,
STORE mode is automatically turned OFF. (PRESET
COUNTER FUNCTION)
• In N RDGS mode, each NS data is stored every time a
trigger signal is generated. STORE mode is in halt status until
the next trigger signal is generated. The STORE mode is automatically turned OFF when memory capacity is full.
In RECALL mode, data are recalled whenever a trigger signal
is generated according to measurement mode. In AUTO
mode, data are automatically recalled at measurement intervals which have been set at the time of recall.
Time Delay
■ NULL FUNCTIONS (NULL)
Y = X – X0
where; X0 : initially set value (NULL value)
Y : computed value
X : measured value
Reset an initially set value (data being displayed) to zero as a
NULL value.
Thereafter a value subtracted by a NULL value is displayed
as a measured value. This null function is used to eliminate
leadwire resistance or cancel initial values.
■ AVERAGING FUNCTION (AVG)
The AVG function is used to average the measured data.
This comes into effect when noise components or variation
of measured data at high-speed sampling must be eliminated
from the measured value.
The calculation expression is given by:
Y = (XAT + XAT-1 + .... + X1)/AT
where; AT : 2 to 100 (number of average data)
X : measured value
Y : computed value
Measured data
Moving average value
Walking Window
• Single Mode (SINGLE)
One datum is sampled at the preset integral time every time
a trigger signal is generated.
Sampling
Trigger
point
• Auto Mode (AUTO)
Data is sampled at a preset integral time and measuring intervals.
1 2
3 4
1st window
2nd window
3rd window
4th window
1
5
10
Sampling Times
Averaging Function (Number of averaging = 10)
15
DIGITAL MULTIMETERS
FUNCTIONS
RECORDERS
WAVEFORM
MEASURING
INSTRUMENTS
DIGITAL
MEASURING
INSTRUMENTS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
■ MATHEMATICAL FUNCTION (MATH)
■ STORING FUNCTION (STORE)
• Scaling
The scaling function serves not only to indicate a multiple
with preset multiplying factors but also to determine deviation from a preset reference value.
Store the measured data in built-in memory or IC card
memory. When an IC memory card is installed, the measured data is stored in the IC card memory.
where ; Y : computed value,
X : measured value,
A and B : constants,
where :
–1,999,999E9 ≤ A ≤ 1,999,999E9
–1,999,999E9 ≤ B ≤ 1,999,999E9, B ≠ 0
• Decibel
The decibel function serves to perform logarithmic operations (common logarithms) for measured values (or null or
averaging values).
Y = C x log10 (X/D),
where ; Y : computed value,
X : measured value,
C and D : constants,
where :
–1,999,999E9 ≤ C ≤ 1,999,999E9
–1,999,999E9 ≤ D ≤ 1,999,999E9, D ≠ 0
• Comparator
The comparator function compares a measured value (or
null or averaging values) with the reference value to determine which value is larger or smaller if the measured value
is within the limits.
X ≥ H ... High
H > X > L ... Pass
L ≥ X ... Low
X : measured value,
H and L : constants,
where :
–1,999,999E9 ≤ H ≤ 1,999,999E9
–1,999,999E9 ≤ L ≤ 1,999,999E9
If H and L are set to the same value, whichever value is
larger or smaller is indicated. And the computed result is
displayed and output to outer computer pin.
The STORE function has three operations depending on
measurement modes as follows:
• In AUTO Mode
In AUTO mode, press the STORE key to store data in
memory. If the memory overflows, the oldest data are automatically erased and replaced by the newest data. If the
STORE mode is OFF, data is no longer stored. In AUTO
mode, if a trigger signal is generated, data can be written to
the memory, leaving NS-1 data before generating a trigger
signal. (PRE-TRIGGER FUNCTION)
Old data
(1st)
Store mode OFF
(51th)
Trigger signal
is generated
(200th)
NS = 150
Pre-trigger (149)
Data No. 0
Data No. –149
Data
No. –1
Data No. 350
Data storing direction
Maximum No. of data (500)
Memory capacity
• Built-in memory:
• 8K bytes
IC card memory:
• 16K bytes
IC card memory:
• 64K bytes
IC card memory:
1,000 data
500 data
1,500 data
8,000 data
Pre-trigger Function
• In SINGLE Mode
In SINGLE mode, data measured in each sampling cycle are
stored every time a trigger signal is generated. If NS data that
was preset are stored, STORE mode is automatically turned
OFF. (PRESET COUNTER FUNCTION)
• In N RDGS Mode
NS data are stored every time a trigger signal is generated.
The STORE mode is in halt status until the next trigger signal
is generated. If the memory overflows, the STORE mode is
automatically turned OFF. (POST TRIGGER FUNCTION)
15
DIGITAL MULTIMETERS
Y = (X – A)/B,
RECORDERS
WAVEFORM
MEASURING
INSTRUMENTS
DIGITAL
MEASURING
INSTRUMENTS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
• In N RDGS Mode
NS data are output at the preset measurement intervals
whenever a trigger signal is generated. Then the mode is in
the halt state. If the trigger key is pressed again, recalling of
data is started. If the final data are output, the recall mode is
automatically turned OFF.
■ IC MEMORY CARD
The 7560 Series are the first DMM with IC memory card
(optional).
IC memory card allows you to:
• Save up to 8,000 measured data.
• Functions, ranges, and mathematical functions ON/
OFF, sample speed and so on, are saved.
• Setting programs using the front panel keys and measured data can be stored.
• Automatic loading of setting program at power ON.
7560 Series get the more applications by the above IC
memory card functions.
The data format of IC memory card is compatible with that
of YOKOGAWA LR Series Recorders, so the measured data
can be recorded in analog form on strip chart.
(DMM)
(Programming)
1. DC V .....
2. DC A .....
3. AC V .....
4. AC A .....
5. OHM .....
Data
• In SINGLE Mode
Data are recalled every time a trigger signal is generated.
When the final data are output, the recall mode is automatically turned OFF.
TRIG
• In AUTO Mode
Data are output at the preset measurement intervals. When
the final data are output, the recall mode is automatically
turned OFF.
• Automatic Loading
IC memory cards allow measurement parameters and conditions to be set without using the panel keys. If a memory
card containing set data is inserted and then the power is
turned ON, the memory contents are automatically read and
programmings are set.
This is very useful for measurements that are repeated under
the same conditions, and this can be a powerful support tool
such as in the production line.
• Programming Function
If an IC memory card is installed, up to 20 steps can be programmed either by front panel key operations or via communication interfaces. The required functions, ranges, and
mathematical functions can be set ON or OFF, and multichannel measurements can be performed in a combination
with scanners.
END
■ RECALL FUNCTION (RECALL)
Recalls the stored data from the memory. The recall functions are as follows:
20. DC A
Scanner
Example of Programming
IC memory card
Connection Diagram for LR Series Recorder
■ CALIBRATION FUNCTION
The multimeters can be calibrated either by front panel key
operations or via communication interfaces. The multimeters are calibrated without opening the case and any special calibration skills.
15
DIGITAL MULTIMETERS
■ AUTO ZERO FUNCTION
AUTO ZERO (AZ) compensates for zero drift which will be
generated in the internal circuit every time sampling is executed. In high-speed sampling mode, measurement time is
decreased by eliminating AUTO ZERO function.
RECORDERS
WAVEFORM
MEASURING
INSTRUMENTS
DIGITAL
MEASURING
INSTRUMENTS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
SELECTION GUIDE
< Selection Symbols >
<7560 Series >
• 6-1/2 digits
• 1,999,999
• DC V accuracy: ±0.003%
7561
7562
GP-IB
GP-IB
GP-IB
SPECIFICATIONS
■ GENERAL SPECIFICATIONS
Operating Principle: Feedback pulse width modulation method.
Sample Mode: Auto/Single/N reading.
Sampling Interval: 3 ms to 60 min (7561 and 7562), (1 ms min., 1s
at 3s or more)
Maximum Reading: 1,999,999
Overrange Information: — oL— sign display.
Data Memory: 1,000 data, measured data can be stored and
recalled: (STORE/RECALL).
Ranging: AUTO, MANUAL, (remote control and programming
possible).
Operating Temperature Range: 5 to 40˚C (41 to 104˚F).
Humidity Range: 20 to 80% relative humidity.
Warmup Time: Approx. 60 minutes to rated accuracy.
Power Requirements: 100 or 115, 200 or 230 V AC (must be
specified), 50 or 60 Hz.
Power Consumption: 20 VA max.
Dimensions (Approx.): 213(W) × 88(H) × 330(D) mm,
(8-3/8 × 3-1/2 × 13").
Weight (Approx.): 3 kg (6.6 lbs).
DIGITAL MULTIMETERS
15
DIGITAL
MEASURING
INSTRUMENTS
WAVEFORM
MEASURING
INSTRUMENTS
RECORDERS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
7560 Series (Common to 7561 and 7562)
■ DC VOLTAGE (DC V)
● Ranges:
*16.7 ms shows 16.66
Integrating Time
(500/200 ms)
Range
Max. Reading
Integrating Time
(100/20/16.7 ms)*
Resolution
Max. Reading
Integrating Time
(2.5/1.2 ms)
Resolution
Max. Reading
Resolution
200 mV
199.9999
0.1 µV
199.999
1 µV
199.99
10 µV
2,000 mV
1,999.999
1 µV
1,999.99
10 µV
1,999.9
100 µV
20 V
19.99999
10 µV
19.9999
100 µV
19.999
1 mV
200 V
199.9999
100 µV
199.999
1 mV
199.99
10 mV
1,000 V
1,100.000
1 mV
1,100.00
10 mV
1,100.0
100 mV
Input
Resistance
Max. Input
>1 GΩ
±1,000 V pk (10s)
±600 V pk continuously
between Hi and Lo
10 MΩ ±1%
±1,000 V pk continuously
● Accuracy (Integrating Time 500 ms): ±(% of reading + digits)
Range
24 hours, 23±1˚C
90 days, 23±5˚C
Temperature Coefficient
(5 to 18, 28 to 40˚C)
1 year, 23±5˚C
200 mV
0.004 + 30(6) {4}
0.006 + 40(8) {4}
0.01 + 40(8) {4}
0.0007 + 5(.6) {.2}
2,000 mV
0.0025 + 10(3) {3}
0.0045 + 15(3) {3}
0.0075 + 15(3) {3}
0.00055 + 1(.2) {.1}
20 V
0.003 + 10(3) {3}
0.005 + 15(3) {3}
0.009 + 15(3) {3}
0.00065 + 1(.2) {.1}
200 V
0.0045 + 10(3) {3}
0.009 + 15(3) {3}
0.016 + 15(3) {3}
0.00075 + 1(.2) {.1}
1,000 V
0.005 + 10(3) {3}
0.01 + 20(3) {3}
0.017 + 20(3) {3}
0.0008 + 1(.2) {.1}
● Accuracy at 24 hours, 23±1˚C is the value for the calibration
standard.
● Auto Zero ON, Null.
●Integrating Time: At 200 ms, 2 is added to the value (digits) in
integrating time 500 ms.
● ( ) indicates the value (digits) in integrating time 100 ms. For
integrating time 20/16.7 ms, 2 is added to the value (digits)
enclosed in the parentheses.
● { } indicates the value (digits) in integrating time 2.5 ms. For
integrating time 1.2 ms, 2 is added to the value (digits) enclosed in
the brackets.
● At Auto Zero OFF, temperature coefficient of ±(0.0015% of range
+ 25 µV)/˚C is added (at 5 to 40˚C).
● Common Mode Rejection: 120 dB or more.
Integrating time 500/200/100/20/16.7 ms, RS = 1 kΩ, 50/60 Hz
±0.1%
● Normal Mode Rejection: 60 dB or more.
Integrating time 500/200/100/20/16.7 ms, 50/60 Hz ±0.1%
● Maximum Allowable Voltage: ±500 V peak between Lo and case.
■ DC CURRENT (DC A)
● Accuracy (Integrating Time 500 ms):
±(% of reading + digits)
Integrating Time
(500/200/100/20/16.7 ms)
Range
Max. Reading
Resolution
Integrating Time
(2.5/1.2 ms)
Max. Reading
Resolution
Input
Resistance
Range
1 year, 23±5˚C
2 mA
0.05 + 100
20 mA
0.05 + 20
2 mA
1.99999
10 nA
1.9999
100 nA
< 110 Ω
20 mA
19.9999
100 nA
19.999
1 µA
< 11 Ω
200 mA
0.05 + 20
2,000 mA
0.1 + 40
200 mA
199.999
1 µA
199.99
10 µA
< 1.2 Ω
2,000 mA
1,999.99
10 µA
1,999.9
100 µA
< 0.3 Ω
● Temperature Coefficient: ±(1/10 of measurement accuracy)/˚C.
● Auto Zero ON.
● Integrating Time: At 200/100/20/16.7 ms, 20 is added to the
value (digits) in integrating time 500 ms.
● Allowable Current: 2 A (2 A fuse installed).
15
DIGITAL MULTIMETERS
● Ranges:
DIGITAL
MEASURING
INSTRUMENTS
WAVEFORM
MEASURING
INSTRUMENTS
RECORDERS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
■ RESISTANCE (OHM)
● Ranges:
Integrating Time
(500/200 ms)
Range
Integrating Time
(100/20/16.7 ms)
Max. Reading
Integrating Time
(2.5/1.2 ms)
Max. Reading
Resolution
Resolution
200 Ω
199.9999
100 µΩ
199.999
1 mΩ
2,000 Ω
1,999.999
1 mΩ
1,999.99
10 mΩ
Resolution
Current
through
Unknown
199.99
10 mΩ
1 mA
1,999.9
100 mΩ
1 mA
Max. Reading
20 kΩ
19.99999
10 mΩ
19.9999
100 mΩ
19.999
1 Ω
100 µA
200 kΩ
199.9999
100 mΩ
199.999
1 Ω
199.99
10 Ω
10 µA
2,000 kΩ
1,999.999
1 Ω
1,999.99
10 Ω
1,999.9
100 Ω
20 MΩ
19.9999
100 Ω
19.9999
100 Ω
19.999
1 kΩ
100 nA
200 MΩ
199.999
199.99
10 kΩ
50 nA
1 kΩ
199.999
1 kΩ
1 µA
● Accuracy (4-wire System, Integrating Time 500 ms): ±(% of reading + digits)
Range
24 hours, 23±1˚C
90 days, 23±5˚C
1 year, 23±5˚C
Temperature
Coefficient
(5 to 18, 28 to 40˚C)
0.0012 + 10(2) {.5}
200 Ω
0.007 + 40(6) {4}
0.012 + 50(7) {4}
0.016 + 50 (7) {4}
2,000 Ω
0.005 + 25(4) {3}
0.01 + 35(6) {3}
0.014 + 35(6) {3}
0.001 + 2(.5) {.1}
20 kΩ
0.005 + 20(3) {3}
0.01 + 30(5) {3}
0.014 + 30(5) {3}
0.001 + 2(.5) {.1}
200 kΩ
0.007 + 20(3) {3}
0.011 + 30(5) {3}
0.015 + 30(5) {3}
2,000 kΩ
20 MΩ
200 MΩ
0.02 + 135(15) {20}
0.03 + 150(20) {30}
0.2 + 30(30)
0.2 + 30(30)
2 + 200(200)
0.001 + 2(.5) {.1}
0.05 + 150(20) {30}
0.004 + 2(.5){.1}
0.2 + 30(30)
2 + 200(200)
0.02 + 1(1)
2 + 200(200)
0.05 + 2(2)
● Accuracy at 24 hours, 23±1˚C is the value for
the calibration standard.
● Auto Zero ON, Null.
● Integrating Time: At 200 ms, 2 is added to the
value (digits) in integrating time 500 ms.
) indicates the value (digits) in integrating
time 100 ms. For integrating time 20/16.7 ms,
2 is added to the value (digits) enclosed in the
parentheses.
● { } indicates the value (digits) in integrating
time 2.5 ms. For integrating time 1.2 ms, 2 is
added to the value (digits) enclosed in the
brackets.
● For 20 and 200 MΩ at sampling interval 400
ms or more. Accuracy is not prescribed in
integrating time 1.2/2.5 ms.
● At Auto Zero OFF, temperature coefficient on
200 Ω ranges is ±(0.013% of range)/˚C, on
other ranges ±(0.003% or range)/˚C is added
(at 5 to 40˚C).
● For 2-wire system, 2 mΩ/˚C is added.
● Excluding the influence of leadwires.
● Open Circuit Voltage: Max 10 V (12.5 V max.
on 200 MΩ range).
● Maximum Input Voltage: ±300V peak or 300
V rms (between Hi and Lo).
● Response Time: 0.4 s or less on 2,000 kΩ/20
MΩ ranges, 5 s or less on 200 MΩ range.
●(
■ AC VOLTAGE (AC V) (7562 only)
● Ranges:
Max. Reading
● Auto Zero ON.
● Integrating Time: At 200/100/20/16.7 ms, 20
Integrating Time
(2.5/1.2 ms)
Resolution
Max. Reading
Resolution
200 mV
199.999
1 µV
199.99
10 µV
2,000 mV
1,999.99
10 µV
1,999.9
100 µV
20 V
19.9999
100 µV
19.999
1 mV
200 V
199.999
1 mV
199.99
10 mV
700 V
700.00
10 mV
700.0
100 mV
Input
Resistance
Max. Input
1 MΩ ±2%
approx.
150 pF
700 V rms
or
±1,000 V pk
between
Hi and Lo
● Accuracy (Integrating Time 500 ms): ±(% of reading + digits), 1 year, 23±5˚C
Range
20 to 30 Hz
30 to 45 Hz
45 Hz
to 10 kHz
10 to 20 kHz
20 to 50 kHz
is added to the value (digits) in integrating
time 500 ms.
● AC Coupling: True rms value.
● Input Voltage: 5 to 100% of range (sine wave).
● Response Time: 400 ms or less (to reading
within ±0.2% of final value).
● Crest Factor: Up to 3 (at full scale, up to 2 at
full scale on 700 V range).
● Temperature Coefficient: ±(1/10 of measurement accuracy)/˚C.
50 to
100 kHz
200 mV
0.9 + 200
0.5 + 200
0.3 + 200
0.3 + 300
0.7 + 500
2 + 500
2,000 mV
0.8 + 100
0.4 + 100
0.15 + 100
0.3 + 200
0.5 + 500
2 + 500
20 V
0.8 + 100
0.4 + 100
0.15 + 100
0.3 + 200
0.5 + 500
2 + 500
200 V
1 + 100
0.4 + 100
0.3 + 100
0.3 + 200
0.7 + 500
3 + 500
700 V
1 + 100
0.4 + 100
0.3 + 100
0.3 + 300
—
—
15
■ AC CURRENT (AC A) (7562 only)
● Ranges:
Range
Integrating Time
(500/200/100/20/16.7 ms)
Max. Reading
Integrating Time
(2.5/1.2 ms)
Resolution
Max. Reading
Resolution
Input
Resistance
(50 Hz)
< 110 Ω
2 mA
1.99999
10 nA
1.9999
100 nA
20 mA
19.9999
100 nA
19.999
1 µA
< 11 Ω
200 mA
199.999
1 µA
199.99
10 µA
< 1.2 Ω
2,000 mA
1,999.99
10 µA
1,999.9
100 µA
< 0.3 Ω
● Accuracy (Integrating Time 500 ms): ±(% of reading + digits), 1 year, 23±5˚C
Range
20 to 30 Hz
30 to 45 Hz
45 Hz to 2 kHz
2 mA
1.4 + 350
0.8 + 250
0.5 + 250
2 to 5 kHz
0.8 + 300
20 mA
1.2 + 300
0.8 + 200
0.5 + 200
0.8 + 200
200 mA
1.2 + 300
0.8 + 200
0.5 + 200
0.8 + 200
2,000 mA
1.5 + 300
1.5 + 200
1.0 + 200
1.5 + 200
● Auto Zero ON.
● Integrating Time: At 200/100/20/16.7 ms, 20
is added to the value (digits) in integrating
time 500 ms.
● AC Coupling: True rms value.
● Input Current: 5 to 100% of range (sine wave).
● Response Time: 400 ms or less (to reading
within ±0.2% of final value).
● Crest Factor: Up to 3.
● Temperature Coefficient: ±(1/10 of measurement accuracy)/˚C.
● Allowable Current: 2 A (2 A fuse installed).
DIGITAL MULTIMETERS
Range
Integrating Time
(500/200/100/20/16.7 ms)
DIGITAL
MEASURING
INSTRUMENTS
WAVEFORM
MEASURING
INSTRUMENTS
RECORDERS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
■ SAMPLING INTERVAL
Measurement Intervals
(in Auto Zero OFF)
Integrating Time
1.2 ms
2.5 ms
16.7 ms
3 ms (333/s)
7 ms (143/s)
8 ms (125/s)
15 ms (66.7/s)
25 ms (40/s)
20 ms
45 ms (22.2/s)
30 ms (33.3/s)
55 ms (18.2/s)
100 ms
110 ms (9.1/s)
215 ms (4.7/s)
*
200 ms
210 ms (4.8/s)
415 ms (2.4/s)
*
500 ms
510 ms (2/s)
1015 ms (1/s)
● RS-232-C Interface
● INTERFACE
● GP-IB Interface
Electrical & Mechanical Specifications: Conforms to IEEE St’d
488-1978.
Interface Function & Identification: SH1, AH1, T5, L4, SR1, RL1,
PP0, DC1, DT1, C0.
Address Mode, address and header ON/OFF can be settable.
Transmission System: Start-stop system.
Data Transfer Rates: 75, 150, 300, 600, 1,200, 2,400, 4,800,
9,600 bps.
Hand Shake Mode, bps rate, number of bits and header ON/OFF
can be settable.
■ STANDARD ACCESSORIES
4.
A1013WD
5.
A1023WD
BS standard
1
Name
Part No.
AVAILABLE MODELS
Description
100 V series
(JIS standard)
115 V series
(UL standard)
200 V series
(VDE standard)
230 V series
(SAA standard)
No.
1.
A1007WD
2.
A1006WD
3.
Power supply cord*
AVG: off, MATH: off, function: DC V in a
fixed range, without communication function, using buffer memory.
● Data of sampling interval shows min.
value (fastest sample rate). Measurement
intervals can be settable more than the
value of left table.
A1009WD
Q’ty
1
1
Model
756101
756102
1
756201
1
Version
756202
Suffix Codes
..........................
Description
DC V, DC A, OHM, (GP-IB)
.......................... 6.5 DC V, DC A, OHM, (RS-232-C)
.......................... digits DC V, DC A, OHM, AC V, AC A (GP-IB)
..........................
DC V, DC A, OHM, AC V, AC A (RS-232-C)
-C .................... Always C
-1 .............. 100 V AC (50 or 60 Hz)
-3 .............. 115 V AC (50 or 60 Hz)
Power
Requirements
6.
Measurement lead
B9280TZ
—
1
7.
Remote connector
A1003JD
—
1
A1105EF
0.2 A 100 V
1
/B ......... JIS standard
A1103EF
0.1 A 200 V
1
/D ......... UL standard
A1092EF
—
1
—
—
1
8.
9.
Fuse*
Fuse
— Instruction manual
* Specified one.
-5 .............. 200 V AC (50 or 60 Hz)
-7 .............. 230 V AC (50 or 60 Hz)
/F ......... VDE standard
Power Cord
/G ......... SAA standard
Optional Feature
/DA
D/A converter output
15
■ OPTIONAL ACCESSORIES
No.
Code
Description
Memory card (8 k bytes)
10. Memory card (16 k bytes)
378901
Setting & measured data
378902
Setting & measured data
Memory card (64 k bytes)
378903
Setting & measured data
Name
B9586NG
Dust cap for memory
card slot
751501
EIA (single mounting)
751502
EIA (double mounting)
751503
JIS (single mounting)
751504
JIS (double mounting)
4-wire resistance
measuring lead
751510
0.6 m
12. Input terminal adapter
751512
—
—
11.
Dummy card
Rack mounting kit
—
Order Q’ty
1 unit
(1 pc./unit)
2 units
(1 pc./unit)
1 unit
(1 pc./unit)
DIGITAL MULTIMETERS
*
● *Sampling mode is AUTO, NULL: off,
Measurement Intervals
(In Auto Zero ON)
WAVEFORM
MEASURING
INSTRUMENTS
RECORDERS
DIGITAL
MEASURING
INSTRUMENTS
INDEX
DIGITAL MULTIMETERS
7561 & 7562
11.
DIMENSIONS
10.
Unit: mm (inch)
7561
Rear Panel
12.
10. IC memory card (378901, 378902 and 378903)
11. 4-wire resistance measuring lead (751510)
12. Input terminal adapter (751512)
ORDERING INFORMATION
When ordering, specify:
1. Instrument name, model and codes.
2. Spares, if required.
7562
Rear Panel
< Rack Mounting >
EIA (Double mounting)