...

1001–1023, www.atmos-meas-tech.net/9/1001/2016/ doi:10.5194/amt-9-1001-2016 © Author(s) 2016. CC Attribution 3.0 License.

by user

on
Category: Documents
12

views

Report

Comments

Transcript

1001–1023, www.atmos-meas-tech.net/9/1001/2016/ doi:10.5194/amt-9-1001-2016 © Author(s) 2016. CC Attribution 3.0 License.
Atmos. Meas. Tech., 9, 1001–1023, 2016
www.atmos-meas-tech.net/9/1001/2016/
doi:10.5194/amt-9-1001-2016
© Author(s) 2016. CC Attribution 3.0 License.
EARLINET instrument intercomparison campaigns:
overview on strategy and results
Ulla Wandinger1 , Volker Freudenthaler2 , Holger Baars1 , Aldo Amodeo3 , Ronny Engelmann1 , Ina Mattis1,a ,
Silke Groß2,b , Gelsomina Pappalardo3 , Aldo Giunta3 , Giuseppe D’Amico3 , Anatoli Chaikovsky4 , Fiodor Osipenko4 ,
Alexander Slesar4 , Doina Nicolae5 , Livio Belegante5 , Camelia Talianu5 , Ilya Serikov6 , Holger Linné6 ,
Friedhelm Jansen6 , Arnoud Apituley7 , Keith M. Wilson7 , Martin de Graaf7,c , Thomas Trickl8 , Helmut Giehl8 ,
Mariana Adam9,d , Adolfo Comerón10 , Constantino Muñoz-Porcar10 , Francesc Rocadenbosch10 , Michaël Sicard10 ,
Sergio Tomás10,e , Diego Lange10,f , Dhiraj Kumar10,g , Manuel Pujadas11 , Francisco Molero11 , Alfonso J. Fernández11 ,
Lucas Alados-Arboledas12,13 , Juan Antonio Bravo-Aranda12,13 , Francisco Navas-Guzmán12,13,h ,
Juan Luis Guerrero-Rascado12,13,14 , María José Granados-Muñoz12,13,i , Jana Preißler14,j , Frank Wagner14,21 ,
Michael Gausa15 , Ivan Grigorov16 , Dimitar Stoyanov16 , Marco Iarlori17 , Vincenco Rizi17 , Nicola Spinelli18,19 ,
Antonella Boselli19,3 , Xuan Wang19,20 , Teresa Lo Feudo18,19,k , Maria Rita Perrone21 , Ferdinando De Tomasi21 , and
Pasquale Burlizzi21
1 Leibniz
Institute for Tropospheric Research, Permoserstr. 15, 04318 Leipzig, Germany
Meteorological Institute, Munich, Germany
3 Consiglio Nazionale delle Ricerche, Istituto di Metodologie per l’Analisi Ambientale,
Potenza, Italy
4 B.I. Stepanov Institute of Physics, National Academy of Sciences of Belarus, Minsk, Belarus
5 National Institute of Research and Development for Optoelectronics, Magurele, Ilfov, Romania
6 Max Planck Institute for Meteorology, Hamburg, Germany
7 Royal Netherlands Meteorological Institute, De Bilt, the Netherlands
8 Institut für Meteorologie und Klimaforschung–Atmosphärische Umweltforschung, Karlsruhe Institute of
Technology, Garmisch-Partenkirchen, Germany
9 European Commission, Joint Research Centre, Institute for Environment and
Sustainability, Ispra, Italy
10 Universitat Politècnica de Catalunya, Barcelona, Spain
11 Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Department of Environment,
Madrid, Spain
12 Department of Applied Physics, Faculty of Sciences, University of Granada, Granada, Spain
13 Andalusian Institute for Earth System Research, Granada, Spain
14 Universidade de Évora, Centro de Geofísica de Évora, Évora, Portugal
15 Alomar, Andøya Rocket Range, Andøya, Norway
16 Institute of Electronics, Bulgarian Academy of Sciences, Sofia, Bulgaria
17 Università degli Studi dell’Aquila, CETEMPS/DSFC, L’Aquila, Italy
18 Dipartimento di Fisica Università Federico II di Napoli, Naples, Italy
19 Consorzio Nazionale Interuniversitario per le Scienze Fisiche della Materia, Naples, Italy
20 Consiglio Nazionale delle Ricerche, Istituto superconduttori, materiali innovativi e dispositivi,
Naples, Italy
21 Università del Salento, Dipartimento di Matematica e Fisica, Lecce, Italy
a now at: Meteorological Observatory Hohenpeißenberg, Deutscher Wetterdienst,
Hohenpeißenberg, Germany
b now at: Deutsches Zentrum für Luft- und Raumfahrt, Institute of Atmospheric Physics,
Oberpfaffenhofen, Germany
c now at: Faculty of Civil Engineering and Geosciences, Delft University of Technology, Delft, the Netherlands
2 Ludwig-Maximilians-Universität,
Published by Copernicus Publications on behalf of the European Geosciences Union.
1002
U. Wandinger et al.: EARLINET intercomparison campaigns
d now
at: Met Office, Exeter, UK
at: Institut d’Estudis Espacials de Catalunya, Barcelona, Spain
f now at: Universidad Católica Boliviana San Pablo, Cochabamba, Bolivia
g now at: Universidad Carlos III de Madrid, Madrid, Spain
h now at: Institute of Applied Physics, University of Bern, Bern, Switzerland
i now at: Table Mountain Facility, Jet Propulsion Laboratory, Wrightwood, USA
j now at: Centre for Climate and Air Pollution Studies, School of Physics, National University of Ireland,
Galway, Ireland
k now at: Consiglio Nazionale delle Ricerche, Istituto di Scienze dell’Atmosfera e del Clima, UOS of
Lamezia Terme, Italy
e now
Correspondence to: Ulla Wandinger ([email protected])
Received: 15 September 2015 – Published in Atmos. Meas. Tech. Discuss.: 9 October 2015
Revised: 12 February 2016 – Accepted: 15 February 2016 – Published: 10 March 2016
Abstract. This paper introduces the recent European Aerosol
Research Lidar Network (EARLINET) quality-assurance efforts at instrument level. Within two dedicated campaigns
and five single-site intercomparison activities, 21 EARLINET systems from 18 EARLINET stations were intercompared between 2009 and 2013. A comprehensive strategy for campaign setup and data evaluation has been established. Eleven systems from nine EARLINET stations
participated in the EARLINET Lidar Intercomparison 2009
(EARLI09). In this campaign, three reference systems were
qualified which served as traveling standards thereafter.
EARLINET systems from nine other stations have been compared against these reference systems since 2009. We present
and discuss comparisons at signal and at product level from
all campaigns for more than 100 individual measurement
channels at the wavelengths of 355, 387, 532, and 607 nm.
It is shown that in most cases, a very good agreement of the
compared systems with the respective reference is obtained.
Mean signal deviations in predefined height ranges are typically below ±2 %. Particle backscatter and extinction coefficients agree within ±2 × 10−4 km−1 sr−1 and ± 0.01 km−1 ,
respectively, in most cases. For systems or channels that
showed larger discrepancies, an in-depth analysis of deficiencies was performed and technical solutions and upgrades
were proposed and realized. The intercomparisons have reinforced confidence in the EARLINET data quality and allowed us to draw conclusions on necessary system improvements for some instruments and to identify major challenges
that need to be tackled in the future.
1
Introduction
The European Aerosol Research Lidar Network (EARLINET) was founded in the year 2000 with the major goal
to establish an aerosol climatology for Europe (Pappalardo
et al., 2014). The network has been continuously growing
Atmos. Meas. Tech., 9, 1001–1023, 2016
and currently consists of 27 stations with about 35 individual lidar systems distributed over 16 European countries. Although all systems are specifically designed for aerosol observations in the troposphere and, partly, the stratosphere,
the network comprises a large variety of individual technical
solutions from small laboratory-based systems to mediumsized portable lidars and large container-based instruments.
Moreover, technical improvements, resulting to a large extent from exchange of expertise within the network, lead to
continuous alterations of the setups. Because of this diversity, the need for a rigorous quality-assurance (QA) program
was very clear right from the start of the EARLINET initiative. Consequently, great effort was put into QA activities at
the instrument and algorithm levels over the years.
In the first phase of EARLINET from 2000 to 2003, when
EARLINET was implemented as a research project supported by the European Commission under the Fifth Framework Programme, QA activities were focussed on intercomparisons of lidar systems (Matthias et al., 2004) and of dataevaluation algorithms (Böckmann et al., 2004; Pappalardo
et al., 2004). In order to check the quality of the instruments
within the network, an intercomparison strategy was developed based on the application of reference lidar systems that
can serve as traveling standards (Matthias et al., 2004). All 19
EARLINET systems, which were part of the network at that
time, had been intercompared, some in dedicated campaigns,
but most of them pairwise by comparison with the mobile
reference systems from the EARLINET stations in Hamburg and Munich. Comparisons were exclusively performed
for the products provided to the EARLINET database, i.e.,
profiles of particle backscatter and extinction coefficients
(Matthias et al., 2004). As a general result, it was found
that typical mean deviations of particle backscatter coefficients were 10 % in the planetary boundary layer (PBL) and
1 × 10−4 km−1 sr−1 in the free troposphere and thus well below the thresholds of 25 % and 5 × 10−4 km−1 sr−1 , respectively, representing the predefined quality criteria. Only few
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
comparisons were made for particle extinction coefficients,
but mean deviations were also small in these cases with values of less than 5 % or 0.01 km−1 .
During EARLINET–ASOS (Advanced Sustainable Observation System), an Integrated Activity within the Sixth
Framework Programme from 2006 to 2011, QA activities
were intensified and included also the development of tools
for internal tests of accuracy and temporal stability of individual lidar systems at any time, i.e., independent of dedicated intercomparisons with reference instruments (Freudenthaler et al., 2016). The QA activities have been continued
in the framework of ACTRIS (Aerosols, Clouds, and Trace
gases Research InfraStructure; funded as Integrated Infrastructure Initiative in the Seventh Framework Programme and
as Integrating Activity in Horizon 2020), part of which EARLINET has been since April 2011.
In this paper, we report on instrument intercomparison
campaigns performed within EARLINET–ASOS and ACTRIS from 2009 to 2013. Focus of the activities was on
the development and test of new reference systems, the integration of new EARLINET stations, and the test of new
or considerably enhanced instruments at initial EARLINET
stations. It should be noted that in the period from 2000 to
2003 the major goal of EARLINET was to provide independent measurements of particle extinction and backscatter coefficients by applying the Raman lidar method at least
at one wavelength, preferably in the UV. Since then, a large
number of EARLINET instruments have been upgraded to
so-called 3 + 2 Raman lidar systems. The term 3 + 2 stands
for the independent measurement of three backscatter coefficients (at 355, 532, and 1064 nm) and two extinction coefficients (at 355 and 532 nm) by the use of an Nd:YAG
laser with frequency doubling and tripling and the detection
of elastic-backscatter signals at the three laser wavelengths
and of vibration-rotation or pure rotational Raman signals of
a reference gas (nitrogen and/or oxygen) at the two shorter
wavelengths. With this measurement capability it is possible to retrieve not only optical but also microphysical particle properties (e.g., Müller et al., 1999; Veselovskii et al.,
2002; Böckmann et al., 2005; Müller et al., 2015). In the first
EARLINET period, 11 out of the 19 EARLINET stations
delivered extinction and backscatter coefficients in the UV,
but only two of them were 3 + 2 systems (Matthias et al.,
2004). Currently (in 2015), there are 22 3 + 2 systems at
18 EARLINET stations, and their number is steadily growing. Many systems have polarization measurement capabilities in addition, i.e., the particle linear depolarization ratio
is measured at least at one wavelength (Freudenthaler et al.,
2009; Belegante et al., 2016; Bravo-Aranda et al., 2016). This
quantity contains information about the presence of large,
nonspherical particles and is an indispensable parameter for
aerosol typing, in particular for the identification of mineral
dust in the atmosphere.
The increased number and complexity of lidar systems
within the network requires also an improved QA strategy.
www.atmos-meas-tech.net/9/1001/2016/
1003
The major challenge of the QA efforts lies in the fact that
absolute calibration techniques for aerosol lidar systems do
not exist and that it is practically impossible to validate
aerosol lidar products by comparison with independent measurements externally, e.g., from balloon-borne in situ observations as it is done in the case of water-vapor or ozone lidars (e.g., Leblanc et al., 2011; Nair et al., 2012). Thus, the
direct intercomparison of collocated instruments is the only
objective and commonly accepted way to assess the overall
performance of individual aerosol lidars. The general goal
of such an intercomparison is to identify principal deficiencies, which may lead to systematic errors of the aerosol lidar
products or unreliable results in specific parts of the profile.
For instance, in the near range, lidar systems may suffer
from electronic saturation effects, uncertain optical overlap
functions, and nonlinear signal distortions. In the far range,
the limited dynamic range of data acquisition, together with
electronic signal perturbance, may hinder appropriate background substraction and Rayleigh calibration. Also, principal
optical misalignments or even system design errors may be
discovered. Therefore, a two-step intercomparison strategy
is now applied for EARLINET, starting with a comparison at
signal level to detect the validity range and the uncertainties
of each individual signal part, followed by the comparison of
aerosol products derived from (partly combined) lidar profiles.
In order to cover the larger number of network stations and
to become more flexible with the intercomparison strategy,
it was decided within EARLINET–ASOS to define several
mobile systems as reference lidars. Two 3 + 2 systems with
polarization capability have been newly developed for this
purpose by the EARLINET groups in Hamburg and Potenza.
It was envisaged to perform, in a first step, a specific intercomparison campaign for the two new and three previously
existing mobile reference systems (from Munich, Maisach,
and Minsk), and to travel with these systems to other EARLINET stations for single-site intercomparisons afterwards.
Fortunate circumstances made it possible that not only the
reference lidars but eleven EARLINET systems from nine
stations participated already in the first campaign, the EARLINET Lidar Intercomparison 2009 (EARLI09) in Leipzig,
Germany, in May 2009. Four more systems could be validated by comparison with one of the reference systems in a
second campaign, the Spanish Lidar Intercomparison 2010
(SPALI10), which took place at Madrid, Spain, in October and November 2010. Finally, single-site intercomparisons were realized at five EARLINET stations with six lidar systems between 2009 and 2013. The strategies developed and applied in these campaigns and their results are
discussed in the following. In Sect. 2 an overview of the campaigns and a description of the involved systems is given. The
measurement and data-processing strategies are outlined in
Sect. 3. Results are discussed based on the comparisons at
signal and at product levels in Sect. 4. Further discussion of
the findings is provided in Sect. 5. Finally, Sect. 6 summaAtmos. Meas. Tech., 9, 1001–1023, 2016
1004
U. Wandinger et al.: EARLINET intercomparison campaigns
rizes the conclusions and gives an outlook on future activities.
2
2.1
Instrument intercomparison campaigns
Overview
Figure 1 gives an overview on the stations involved in the
EARLINET intercomparison campaigns between 2009 and
2013. Mobile lidars from the EARLINET stations in Hamburg, Potenza, Munich, Maisach, Bucharest, Cabauw, Minsk,
Ispra, and Garmisch-Partenkirchen were moved to Leipzig
and intercompared during EARLI09 in May 2009, together
with a stationary and a mobile system of the Leipzig site. Afterwards, the reference lidar from Hamburg was brought to
the EARLINET station at Andenes, Norway, for a single-site
intercomparison in October/November 2009. The Munich
system traveled to Sofia to intercompare two lidars at this
site in October 2010. In October/November 2010, the reference lidar from Potenza participated in the SPALI10 campaign in Madrid, where the intercomparison of the systems
from the stations in Évora, Barcelona, Granada, and Madrid
took place. The L’Aquila lidar was intercompared with the
Munich reference system in September 2012. Finally, in October 2013, the Potenza reference lidar was moved to Naples
and Lecce for single-site intercomparisons.
In the following, we introduce the individual campaigns
and the participating instruments in more detail. Table 1 lists
the involved lidar systems by name and institution. Their
measurement channels are detailed in Table 2. The lidar ID
used here includes the ID of the EARLINET station to which
the instrument belongs (two letters) and a number in order to
distinguish systems from stations with more than one instrument.
2.2
EARLI09–EARLINET Lidar Intercomparison
2009
EARLI09 took place at the Leibniz Institute for Tropospheric Research (TROPOS) in Leipzig, Germany, between
4 and 31 May 2009. As mentioned, this campaign was
planned to compare the reference lidar systems, but several other partners took the opportunity to join, and finally 11 EARLINET lidar systems were collocated. In this
way, the campaign became the largest lidar intercomparison
performed so far, with challenging logistical requirements.
Three container-based systems, one van, two trailers, and
three stand-alone systems were placed next to the stationary
Leipzig EARLINET lidar, around and on top of the institute’s
building, and supplied with more than 120 kW of electrical
power and internet connection. The campaign also served
for the implementation and test of the new EARLINET intercomparison strategy (see Sect. 3 for details). Thus, the
first week of the campaign was scheduled for preparations of
hardware and software. Between 11 and 28 May, 20 measureAtmos. Meas. Tech., 9, 1001–1023, 2016
Figure 1. Map of EARLINET and stations involved in the intercomparison campaigns (station IDs: an – Andenes, at – Athens, ba –
Barcelona, be – Belsk, bu – Bucharest, ca – Cabauw, cl – ClermontFerrand, co – Cork, ev – Évora, gp – Garmisch-Partenkirchen, gr
– Granada, hh – Hamburg, is – Ispra, ku – Kuopio, la – L’Aquila,
lc – Lecce, le – Leipzig, lm – Limassol, ma – Madrid, ms/mu –
Maisach/Munich, mi – Minsk, na – Naples, pa – Payerne, pl –
Palaiseau, po – Potenza, sf – Sofia, th – Thessaloniki). Red colors
show stations operating reference systems. Participation of instruments from stations in EARLI09 (yellow), SPALI10 (green), and
single-site intercomparisons (blue) is indicated. Black dots represent stations which were not involved in the 2009–2013 intercomparisons.
ment sessions of 1–3 h duration on 11 days were performed.
Radiosondes were launched for each session. Daily briefings,
including an expert’s weather forecast, served for the planning of the sessions and for the discussion of results from
the previous day. The latter were obtained using a new dataevaluation concept (see Sect. 3) that allowed us to handle, on
a daily basis, the signals from more than 100 lidar channels
provided by the 11 lidar systems which are described in the
following (see also Tables 1 and 2).
The Atmospheric Raman Lidar (ARL2-mobile, ID: hh01)
of the Max Planck Institute for Meteorology (MPI-MET) in
Hamburg is a multiwavelength Raman lidar. With 26 measurement channels (see Table 2) it is the most extensive EARLINET lidar. The emitter is a 440 mJ Nd:YAG laser (Quantel, Brilliant B). The system has two unique features. Firstly,
it covers the altitude range from about 50 m above ground
up to the stratosphere by applying three separate receivers,
which are fiber-coupled to two Newtonian telescopes with
diameters of 380 (far range) and 150 mm (near range) and a
lens telescope with a diameter of 22 mm (lowest heights), rewww.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
Table 1. EARLINET systems participating in the intercomparison
campaigns
Lidar ID
Lidar name and institution
EARLI09
hh01
ms01
mu01
po01
mi01
le01
le02
is01
bu01
gp01
ca01
ARL2-mobile, MPI-MET, Hamburg, Germany
MULIS, LMU Munich, Germany
POLIS, LMU Munich, Germany
MUSA, CNR-IMAA, Potenza, Italy
LMR-mobile, BISIP, Minsk, Belarus
MARTHA, TROPOS, Leipzig, Germany
PollyXT , TROPOS, Leipzig, Germany
CAML, JRC, Ispra, Italy
RALI, INOE 2000, Bucharest, Romania
HSRL, IMK-IFU, Garmisch-Partenkirchen, Germany
CAELI, KNMI, De Bilt, the Netherlands
SPALI10 (reference system: po01)
gr01
Raymetrics LR331–D400, CEAMA, Universidad de Granada, Spain
ev01
PAOLI, Universidade de Évora, Portugal
ma01
LIDAR-CIEMAT, CIEMAT, Madrid, Spain
ba02
UPC-MRL, UPC, Barcelona, Spain
ALI09 (reference system: hh01)
an01
Tropospheric Lidar, Alomar, Andøya Rocket Range, Norway
SOLI10 (reference system: mu01, upgraded)
sf01
Lidar with CuBr laser, IE-BAS, Sofia, Bulgaria
sf02
Lidar with Nd:YAG laser, IE-BAS, Sofia, Bulgaria
LALI12 (reference system: mu01, upgraded)
la01
UV lidar, Università degli Studi dell’Aquila, Italy
NALI13 (reference system: po01)
na01
MALIA, CNISM, Naples, Italy
LELI13 (reference system: po01)
lc01
UNILE lidar, Università del Salento, Lecce, Italy
spectively. Depolarization measurements at 532 nm are utilized with two detection channels, which are directly coupled to another 200 mm Newtonian telescope. The second
remarkable feature of the system is its capability to detect
rotational Raman signals at both 355 and 532 nm with a specific grating technique. In addition, the vibration-rotation signals at 387 nm (nitrogen) and 407 nm (water vapor) are measured. Rotational Raman signals serve for temperature measurements, but can also be used for extinction-coefficient retrievals. Signals are detected by Hamamatsu photomultiplier
tubes (PMTs) in photon-counting detection mode in the UV
and visible wavelength ranges and by Licel/EG&G avalanche
photodiodes (APDs) in analog detection mode at 1064 nm.
The Meteorological Institute of the Ludwig-MaximiliansUniversität (LMU) in Munich participated with two instruments, which both had already served as reference systems in EARLINET. POLIS (Portable Lidar System, ID:
mu01) is a small, rugged lidar system with an exchangeable detector unit. It applies a 50 mJ laser (Big Sky, Ultra
GRM) and a 200 mm Dall-Kirkham Cassegrain telescope.
During EARLI09 the instrument was operated as a twochannel 355 nm system, which detected either parallel- and
cross-polarized elastic backscatter signals or the total elastic
backscattering together with the 387 nm nitrogen Raman signal with Licel/Hamamatsu PMTs for combined analog and
photon-counting detection (Freudenthaler et al., 2009). POLIS was upgraded to three channels in 2010 (see below) and
www.atmos-meas-tech.net/9/1001/2016/
1005
to six channels in 2013 (Freudenthaler et al., 2015). The second system, MULIS (Multichannel Lidar System, ID: ms01),
is a 3 + 2 Raman lidar with polarization measurement capability at 532 nm (Freudenthaler et al., 2009). This lidar performs the EARLINET observations at the station of Maisach,
near Munich. The instrument applies a 1.6 J Nd:YAG laser
(Continuum, Surelite II) and a 300 mm Cassegrain telescope.
Hamamatsu PMTs (UV and visible wavelength range) and a
Licel/EG&G APD (at 1064 nm) are used as detectors. All
elastic-backscatter signals are measured in analog detection
mode. For Raman signals, the combined analog and photoncounting technique (Licel) is applied. MULIS was developed as a reference lidar for the first phase of EARLINET
(2000–2003) and served as a prototype for MUSA (Multiwavelength System for Aerosol, ID: po01), the reference system of CNR–IMAA (Consiglio Nazionale delle Ricerche–
Istituto di Metodologie per l’Analisi Ambientale) in Potenza.
Thus, MUSA has very similar specifications as MULIS, with
the exception that it applies the Licel combined analog and
photon-counting detectors also for the elastic-backscatter
signals at 355 and 532 nm (Madonna et al., 2011).
The fifth EARLINET reference system is the LMRmobile (Lidar Multiwavelength Raman, ID: mi01) of the
B. I. Stepanov Institute of Physics (BISIP), Minsk, Belarus.
It is a compact, scanning, stand-alone system, which applies
a 250 mJ Nd:YAG laser (SOLAR TII, LF-114) and a 300 mm
Cassegrain telescope. The system has six measurement channels in 3 + 2 configuration with polarization discrimination
at 532 nm. Analog detection with PMTs (355, 532 nm) and
an APD (1064 nm) is applied for elastic-backscatter signals.
Photon-counting PMTs are used for Raman signals (387,
607 nm).
MARTHA (Multiwavelength Atmospheric Raman Lidar
for Temperature, Humidity, and Aerosol Profiling, ID: le01)
of TROPOS is the stationary EARLINET lidar at Leipzig,
Germany. It works with a 1.6 J Nd:YAG laser (Spectra
Physics, Quanta-Ray PRO 290) and an 800 mm Cassegrain
telescope. It allows 3 + 2 Raman lidar observations and depolarization measurements at 532 nm. In addition, the instrument has channels for rotational Raman observations at 355
(since 2011, not during EARLI09) and 532 nm, water-vapor
measurements, and dual-field-of-view observations (Mattis
et al., 2004; Schmidt et al., 2013). PMTs in photon-counting
mode are employed in all channels, including at 1064 nm. For
routine, automatic observations, the PollyXT lidar (Portable
Aerosol Raman Lidar System, extended version, ID: le02) is
applied in Leipzig as well. It has a 3 + 2 Raman lidar setup,
utilizing a 450 mJ Nd:YAG laser (Continuum, Inlite III) and
a 300 mm Newton telescope (Althausen et al., 2009). Hamamatsu photon-counting-only PMTs are deployed in all channels. Total and cross-polarized backscattered radiation was
detected at 355 nm during EARLI09 (at 532 nm since the end
of 2011).
CAML (Cloud and Aerosol Micro Lidar, ID: is01) of the
Joint Research Centre (JRC), Ispra, Italy, is a commercial
Atmos. Meas. Tech., 9, 1001–1023, 2016
1006
U. Wandinger et al.: EARLINET intercomparison campaigns
Table 2. Overview of measurement channels of EARLINET systems participating in the intercomparison campaigns; lidar IDs are as in
Table 1. Numbers indicate detection wavelengths; t – total signal, c – cross-polarized signal, p – parallel-polarized signal, RR – rotational
Raman signal, RY – HSRL Rayleigh signal, far – far-range receiver, near – near-range receiver, low – low-range receiver, pol – receiver for
polarization measurements, a – analog detection, p – photon-counting detection, a+p – combined acquisition channels (Licel).
Lidar
Rec.
355t
hh01
far
near
low
pol
p
p
p
ms01
mu01
po01
mi01
le01
le02
is01
bu01
gp01
ca01
a
(a+p)a
a+p
a
p
p
far
near
pol
355c
(a+p)a
355p
(a+p)a
p
a+p
a
a+p
a+p
355RR
387
407
532t
2p
2p
2p
p
p
p
p
p
p
p
a+p
(a+p)a
a+p
p
p
p
a+p
a+p
a+p
gr01
ev01
ma01
ba02
a+p
p
a
a+p
p
p
a+p
a+p
an01
a+p
a+p
p
a
p
a, p
p
p
p
p
532p
p
a
p
a
a+p
a+p
a
p
a+p
p
a
a+p
532RY
532RR
607
2p
2p
2p
2p
a+p
a
a+p
a+p
p
a+p
532c
a
a
a+p
a
a+p
p
p
p
a
a
p
p
a+p
a+p
a+p
a
a
a
a
a
a+p
a+p
a+p
p
a+p
p
p
a+p
a+p
a
p
a
a
a+p
a+p
a+p
a
pb
a
sf01
sf02
la01
c
p
pc
pc
na01
a, p
p, p
p
lc01
a+p
a+p
p
a+p
a
a, p
a+p
1064
a, p
p
a+p
a
a alternative configurations; see text for details
b CuBr laser, emission wavelength at 510 nm
c XeF excimer laser, emission wavelength at 351 nm, Raman-shifted wavelengths at 382 and 403 nm
micropulse lidar supplied by Cimel Electronique. The automatic stand-alone system uses an 8 µJ, 4.7 kHz Nd:YAG laser
and a 200 mm telescope, and it measures 532 nm elasticbackscatter light with a photon-counting APD (Barnaba
et al., 2010).
RALI (Raman Aerosol Lidar, ID: bu01) of the National
Institute of Research and Development of Optoelectronics,
INOE 2000, Bucharest, Romania, is a commercial 3 + 2 Raman lidar from Raymetrics (LR331–D400), including polarization discrimination at 532 nm and a water-vapor detection channel at 407 nm. It applies a 330 mJ laser (Big Sky,
CFR400-10) and a 400 mm Cassegrain telescope. The detection channels are based on Licel/Hamamatsu PMTs for
the UV and visible channels and on a Licel/EG&G APD at
1064 nm (Nemuc et al., 2013; Belegante et al., 2014).
IMK-IFU (Institut für Meteorologie und Klimaforschung–
Atmosphärische Umweltforschung, Karlsruhe Institute of
Atmos. Meas. Tech., 9, 1001–1023, 2016
Technology) participated in EARLI09 with a newly developed 532 nm High Spectral Resolution Lidar (HSRL, ID:
gp01). The 3+1 lidar (elastic-backscatter signals at 355,
532, 1064 nm and Rayleigh signal at 532 nm) applies a 0.5 J
Nd:YAG laser (Quanta Ray, LAB-150-30) and a 300 mm
Cassegrain telescope. The Rayleigh signal in the second
532 nm channel is separated with an iodine filter. Analog
signal detection with actively stabilized Hamamatsu 7400-03
PMTs and a pin photodiode at 1064 nm (both from Romanski
Sensors) is utilized.
CAELI, the CESAR (Cabauw Experimental Site for Atmospheric Research) Water Vapor, Aerosol, and Cloud Lidar
(ID: ca01), was developed by the National Institute for Public
Health and the Environment (RIVM), Bilthoven, the Netherlands, and is now operated by the Royal Netherlands Meteorological Institute (KNMI), De Bilt, the Netherlands (Apituley et al., 2009). CAELI works with a 1.6 J Nd:YAG laser
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
(Continuum, PowerLite Precision II 9030 SI) and has two
3 + 2 setups with a water-vapor Raman channel, one coupled
to a 150 mm Newton telescope for near-range measurements
and one to a 570 mm Newton telescope for far-range observations. In addition, a 50 mm lens telescope is used to measure
parallel- and cross-polarized 532 nm signals. Licel data acquisition technique with Hamamatsu PMTs for the UV and
visible wavelength range and EG&G APDs for 1064 nm is
applied in all channels.
2.3
SPALI10 – Spanish Lidar Intercomparison 2010
The second dedicated intercomparison campaign brought together the EARLINET systems of the Iberian peninsula from
the stations in Évora, Barcelona, Madrid, and Granada. Comparisons were made against the reference system MUSA
from CNR–IMAA in Potenza (ID: po01), which was successfully tested in EARLI09 before. The campaign called
SPALI10 took place at the Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas (CIEMAT), Department of Environment, Atmospheric Pollution Division,
in Madrid, Spain, between 18 October and 5 November 2010.
The campaign strategy followed the rules established in
EARLI09. The first week of the campaign was used for instrument setup and tests of the automated preprocessing of
data (see Sect. 3). During the following 2 weeks, measurement sessions were regularly scheduled during day and night.
All in all, 29 sessions of 1–3 h duration were performed.
Radiosondes were launched systematically during the whole
field campaign for each measurement session.
All systems of the SPALI10 campaign are multiwavelength Raman lidars (see Tables 1 and 2). The Granada group
operates a Raymetrics LR331–D400 system (ID: gr01) with
specification as described for the Bucharest system above
(Guerrero-Rascado et al., 2008, 2009). PAOLI (Portable
Aerosol and Cloud Lidar, ID: ev01) from Évora is a 3 + 2
system of PollyXT type (Althausen et al., 2009) with a 450 mJ
Nd:YAG laser (Continuum, Inlite III) and a 300 mm Newton telescope. Hamamatsu photon-counting-only PMTs are
applied in all channels. Cross-polarized backscattered radiation, together with a total signal, is detected at 532 nm. The
LIDAR-CIEMAT system (ID: ma01) from Madrid and the
UPC MRL (Universitat Politècnica de Catalunya Multispectral elastic-Raman Lidar, ID: ba02) from Barcelona are both
3 + 2 systems without polarization discrimination, but with
a water-vapor channel in the case of UPC MRL. LIDARCIEMAT makes use of a 1 J Spectra Physics laser (LAB-17030) and a 300 mm Newtonian telescope. UPC MRL applies
a 365 mJ laser (Quantel, Brilliant) and a 355 mm Schmidt–
Cassegrain telescope (Kumar et al., 2011). The detection
channels of both systems are based on the Licel/Hamamatsu
PMT and Licel/EG&G APD acquisition systems.
www.atmos-meas-tech.net/9/1001/2016/
1007
2.4
Single-site intercomparison campaigns
Further intercomparisons were performed by moving one of
the reference systems to specific EARLINET sites. Actually,
this is the basic strategy applied for EARLINET intercomparisons at instrument level. It is planned to continue these
kinds of intercomparisons over the years in order to validate
each EARLINET system with a reference system from time
to time. Nevertheless, respective efforts are large and require
appropriate funding. Five activities were carried out between
2009 and 2013.
From 22 October to 5 November 2009 the reference system ARL2-mobile of MPI Hamburg (ID: hh01) was stationed
at Andøya, Norway, in order to compare the Alomar Tropospheric Lidar (ID: al01). During ALI09 (Alomar Lidar Intercomparison 2009) simultaneous measurements were performed on nine days. On two days (4 and 5 November) radiosondes were launched to support the signal calibration.
The Alomar Tropospheric Lidar is a 3 + 2 multiwavelength
Raman lidar with a 1 J laser (Spectra Physics, GCR-6-30)
and a 175 mm Newtonian telescope. The data acquisition is
based on the Licel/Hamamatsu PMT and Licel/EG&G APD
concepts. The optical receiver of the Alomar Tropospheric
Lidar had been changed considerably in the time before the
intercomparison, and the campaign was also used to fix remaining technical issues.
The lidar intercomparison SOLI10 (Sofia Lidar Intercomparison 2010) took place at the Institute of Electronics, Bulgarian Academy of Sciences (IE-BAS), in Sofia, Bulgaria,
between 9 and 14 October 2010. POLIS from LMU Munich (ID: mu01) served as the reference system. POLIS had
been upgraded to three channels in summer 2010, including
355 nm with polarization discrimination and either 532 nm
total or 387 nm, and had been intercompared with the reference lidar system MULIS (ID: ms01) in Maisach again. POLIS was transported to Sofia to intercompare both lidar systems of IE-BAS, one working with a 0.1 mJ CuBr vapor laser
at 510 nm, and the other with a 1 J Nd:YAG laser (EKSMA)
at 532 and 1064 nm (Stoyanov et al., 2011). Both systems are
elastic-backscatter lidars. The CuBr system (ID: sf01) uses a
150 mm Cassegrain telescope and a photon-counting PMT
as the detector. The Nd:YAG system (ID: sf02) applies a
350 mm Cassegrain telescope and analog detection. The latter system points out of a lab window under 58◦ zenith angle. Thus, the intercomparisons were made separately for the
two systems, using the respective scan angle for the POLIS
measurements.
The L’Aquila Lidar Intercomparison 2012, LALI12, was
performed at the EARLINET site of the Dipartimento di
Fisica, Università degli Studi dell’Aquila, in L’Aquila, Italy,
between 10 and 15 September 2012. One daytime and three
nighttime sessions covering one 60 min and six 30 min intercomparison periods were carried out. Also here, POLIS (ID:
mu01) served as the reference system. The lidar at L’Aquila
(ID: la01) is a UV aerosol and water-vapor lidar, which apAtmos. Meas. Tech., 9, 1001–1023, 2016
1008
plies a XeF excimer laser (Lambda Physik, EMG 150 MSC),
a 200 mm telescope, and PMTs in photon-counting mode
(Rizi et al., 2004). The emission wavelength is only slightly
different from the third harmonic of a Nd:YAG laser, and
thus the wavelength shift of the received elastic-backscatter
(351 nm) and nitrogen Raman signals (382 nm) is neglected
in the comparisons.
The lidar system MALIA (Multiwavelength Aerosol Lidar
Apparatus, ID: na01) of the Consorzio Nazionale Interuniversitario per la Scienze Fisiche della Materia (CNISM) in
Naples, Italy, was intercompared with the Potenza reference
lidar MUSA (ID: po01) during the Naples Lidar Intercomparison 2013, NALI13, from 14 to 18 October 2013. Two
daytime and three nighttime measurement periods of 30 min
to 4 h were covered. MALIA is a 10-channel system based
on a 0.5 J Nd:YAG laser (Quantel, Brilliant-B) and a 0.3 m
Newtonian telescope. Signals at 355 nm (total) and 532 nm
(cross- and parallel-polarized) are detected with both photoncounting and analog channels. The Raman return at 387 nm is
split to enter a high-signal and a low-signal photon-counting
channel. Further photon-counting channels detect the Raman signals at 407 and 607 nm. Data acquisition is based on
150 MHz photon counters and 12 bit analog–digital converters.
From 21 to 25 October 2013, the Lecce Lidar Intercomparison LELI13 took place at the Università del Salento in
Lecce, Italy. Again, the MUSA lidar (ID: po01) served as the
reference system. Four daytime and five nighttime sessions
were performed. The EARLINET station of Lecce operates
a multiwavelength Raman lidar (UNILE lidar, ID: lc01) with
a 1.4 J Nd:YAG laser (Quantel, YG981E) and a 0.3 m Newtonian telescope (Perrone et al., 2014). The 3 + 2 system has
polarization discrimination at 355 nm and a water-vapor Raman channel at 407 nm. Licel data acquisition technique with
Hamamatsu PMTs for the UV and visible wavelength range
and an EG&G APD for 1064 nm is applied in all channels.
3
Measurement and data-processing strategies
The participation of a relatively large number of lidar systems in an intercomparison campaign, like EARLI09 and
SPALI10, requires the development and application of coordinated observation and data-evaluation strategies. For instance, it is necessary to have preliminary comparison results at hand as soon as possible after each measurement session in order to detect and remove system faults immediately.
Particular attention must be paid in the beginning of a campaign when systems had been moved before. Specific care is
also necessary when systems are brand new, as was the case
in EARLI09 for the new reference systems. In addition, introducing differences in the comparisons by using different
analysis software should be avoided. These considerations
led to the development of a special version of the Single
Calculus Chain (SCC, D’Amico et al., 2015, 2016) before
Atmos. Meas. Tech., 9, 1001–1023, 2016
U. Wandinger et al.: EARLINET intercomparison campaigns
EARLI09 in order to preprocess the raw lidar data in a common way instantaneously. An additional piece of software,
developed at LMU Munich, served for the direct comparison
at signal level, i.e., necessary interpolation, smoothing, and
weighting as well as visualization of signals and determination of signal deviations. Finally, a modified version of the
SCC optical products module (Mattis et al., 2016) was used
to calculate particle extinction and backscatter coefficients
from the processed signals in order to perform comparisons
at product level. The respective concepts are outlined in the
following.
In all intercomparison campaigns the lidar systems were
collocated on a flat terrain within about 100 m distance. The
laser beams pointed close to the zenith (except sf02; see
above), which made it very likely that all instruments measured the same atmospheric volume within the averaging
time. Several sessions were scheduled for every day of the
campaigns (weather permitting), one at daytime and one at
night if possible. Each session lasted several hours with the
goal to find at least a 30 min period in each session with stable atmospheric conditions and with all lidar systems up and
running. In order to be as flexible as possible in the selection
of final comparison periods, the raw signals were stored with
1 min resolution. The complete data sets of these raw signals
from all systems had to be delivered without any preprocessing to a common database server shortly after each session.
The raw-signal formats had been predefined, following
standards set for the EARLINET SCC. Each data set includes
a header with all information necessary for further processing
of the signals. Some basic, fixed parameters of each system
had been collected in a system database. Using the header
and database information, all signals were then preprocessed
by the modified version of the SCC. The preprocessor performs trigger-delay shift, dead-time correction, background
subtraction, and range correction. If requested, the preprocessor also combines near-range and far-range signals, photoncounting and analog signals (gluing), and parallel- and crosspolarized signals into a total profile using given calibration
ranges or values. After this individual signal preprocessing
and after selection of an appropriate comparison period, the
signals were averaged, typically over 30 to 120 min, in order
to improve the signal-to-noise ratio.
Figure 2 illustrates the processing steps at signal level for
the example of 387 nm signals measured with nine systems
in eleven channels during EARLI09 on 25 May 2009, 21:00–
23:00 UTC. The channels are distinguished by color, and the
legend provides the system ID (see Table 1) as well as a
three-digit channel ID with the following meaning:
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
f__
n__
l__
x__
–
–
–
–
signal from far-range telescope
signal from near-range telescope
signal from low-range telescope
signal from a system with one telescope
_t_
_p_
_c_
_s_
–
–
–
–
total signal
parallel-polarized signal
cross-polarized signal
sum of parallel- and cross-polarized signals
__a
__p
__g
–
–
–
analog signal
photon-counting signal
analog and photon-counting glued signal (Licel).
In Fig. 2a, the individual signals are shown after preprocessing with the SCC. Here, the averaged output signals provided by the SCC preprocessor still have the original range
resolutions from 3.75 to 60 m. In addition, a range offset
may occur because of different lidar location altitudes above
ground (e.g., when a lidar is operated in a building or on top
of a building and compared against a reference system in a
van or container at ground level). Furthermore, pointing angles of the systems are typically between 0 and 5◦ and require
further altitude corrections. In order to allow for a point-bypoint comparison, the signals were re-binned to a common
height resolution of 60 m and to common height levels considering the individual system altitudes and the lidar zenith
angles. The signal noise at higher altitudes was reduced by
further stepwise progressive smoothing with up to 960 m resolution. The result is presented in Fig. 2b. In order to compare the signals quantitatively, they were normalized in the
height range between 3.5 and 6.5 km, where the deviations
are small and the signal-to-noise ratios are high.
Usually, comparisons should be made against a reference
system for all individual wavelengths and polarization states.
However, in EARLI09, none of the reference systems were
considered to be proven already. Therefore, the chosen strategy was to construct a mean signal, or common reference,
from the best parts of all available signals. Ideally, this common reference should be close to the unknown true signal.
For this purpose, range-dependent weights are assigned to
the individual signals by an expert’s guess reflecting an assumed accuracy; see Fig. 2c. A weight of zero means that
the respective part of the signal, e.g., the range of incomplete overlap, is omitted. A weight of 1 is assigned to ranges
that appear trustworthy. Then, a weighted mean signal is
calculated as a first guess of the common reference. Afterwards, the expert’s weights are successively decreased by a
factor commensurate with the range-dependent signal deviation from the first-guess mean signal; see Fig. 2d. In this
way, highest weights are assigned to the best signal parts, and
the final common reference is calculated. In the stratosphere,
where an aerosol-free range can be assumed, the mean signal is replaced by a calculated signal from actual radiosonde
www.atmos-meas-tech.net/9/1001/2016/
1009
data (pure molecular Rayleigh or Raman signal), fitted to
the common reference at an appropriate height (usually at
about 15 km). The radiosonde data were taken from local radiosonde ascents during the experiment.
The approach of a common reference was applied in
EARLI09 only. In all other campaigns, the reference system
was considered as the standard to which the other systems
were compared. Point-by-point deviations as well as mean
deviations in certain height ranges are used to assess the quality of the signals.
If Pref (zi , λ) is the reference signal at wavelength λ (either
the common reference or the signal from the reference system), the relative deviation of an individual signal P (zi , λ)
from this reference signal is calculated for each individual
height zi (to which the signals were commonly binned) as
1P (zi , λ) =
P (zi , λ) − Pref (zi , λ)
.
Pref (zi , λ)
(1)
The relative deviations are shown in Fig. 2e for the example
case of 25 May 2009.
The mean relative systematic deviation (relative bias) of
an individual signal from the reference signal over a height
range 1z = zL − zK , i.e., L − K + 1 height bins, is defined
as
L
P
1P (1z, λ) =
1P (zi , λ)
i=K
.
L−K +1
(2)
The mean relative systematic deviation is used to assess the
quality of signals in certain atmospheric height ranges (e.g.,
boundary layer, free troposphere, stratosphere).
For the comparison at product level, aerosol optical parameters were computed using a special version of the SCC optical products module (Mattis et al., 2016). This version is able
to treat the preprocessed, re-binned, and normalized signals,
and also the common reference, on the common height grid
(with 60 m vertical resolution in EARLI09). Thus, point-bypoint comparisons and the calculation of mean deviations are
possible for the products in the same way as for the signals.
We use the absolute deviation,
1c(zi , λ) = c(zi , λ) − cref (zi , λ),
(3)
of a coefficient c (either extinction or backscatter coefficient)
from the reference coefficient cref at individual heights and
the mean absolute systematic deviation (absolute bias) in certain height ranges,
L
P
1c(1z, λ) =
1c(zi , λ)
i=K
L−K +1
,
(4)
to investigate the quality of optical products.
Atmos. Meas. Tech., 9, 1001–1023, 2016
1010
U. Wandinger et al.: EARLINET intercomparison campaigns
(a)
(c)
11
10
2E13
9
8
Initial signal weight
Range-corrected signal, a.u.
1E13
1E12
1E11
7
6
5
4
3
2
1
0
0
1E10
1
2
3
4
5
6
7
8
9
10 11 12
14
16
18
20
14
16
18
20
14
16
18
20
Height above ground, km
(d)
1E9
11
10
9
8
0
1
2
3
4
5
6
7
8
9
10 11 12
14
16
18
Final signal weight
2E8
20
Height above ground, km
(b)
6
5
4
3
6
2
Common reference
po01-xtg
mu01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
hh01-ntp
hh01-ftp
bu01-xtg
ca01-ntg
ca01-ftg
rs09052503
1
1
0
0
(e)
0.1
0.03
0
1
2
3
4
5
6
7
8
9
10 11 12
14
16
Height above ground, km
18
1
2
3
4
5
6
7
8
9
10 11 12
Height above ground, km
20
0.5
Relative deviation from common reference
Range-corrected signal, a.u.
7
0.4
0.3
0.2
0.1
0.0
–0.1
–0.2
–0.3
–0.4
–0.5
0
1
2
3
4
5
6
7
8
9
10 11 12
Height above ground, km
Figure 2. Illustration of signal processing for comparison purposes. The measurement was taken during EARLI09 on 25 May 2009, 21:00–
23:00 UTC. (a) Range-corrected signals at 387 nm with individual range resolutions (3.75 to 60 m). (b) Range-corrected lidar signals at
387 nm binned to common height resolution (60 m) and to common height levels, progressively smoothed (60 m up to 3 km, 120 m from
3–6 km, 240 m from 6–9 km, 480 m from 9–12 km, 960 m above), and normalized between 3.5 and 6.5 km. The thick gray line represents the
common reference. A pure molecular signal at 387 nm calculated from radiosonde data (rs09052503) is fitted to the common reference at
10.3 km (shown below 12 km) and, additionally, at 15.3 km (shown above 12 km). (c) Initial weights assigned to the signals for calculation
of a weighted mean signal. (d) Final weights assigned to the signals for calculation of the common reference. (e) Relative deviations of
individual signals from the common reference. For the sake of conspicuity the weights in panels (c) and (d) are successively shifted by a
value of 1 along the y axis.
4
Results
In the following, we present comparison results at signal and
product level. We focus on signals at the wavelengths of 355
(total), 387, 532 (total, parallel-, and cross-polarized), and
607 nm and respective aerosol products, i.e., particle extinction and backscatter coefficients at 355 and 532 nm. We do
Atmos. Meas. Tech., 9, 1001–1023, 2016
not discuss observations at 1064 nm, since there is a separate paper on technical solutions, calibration issues, and
intercomparison results for the infrared wavelength in this
special issue by Engelmann et al. (2016). Furthermore, we
do not show results at product level for the particle depolarization ratio. Depolarization ratio measurements require
specific calibration procedures, which are discussed in dewww.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
tail in this special issue by Bravo-Aranda et al. (2016) and
Freudenthaler (2016). Rotational Raman lidar signals at 355
and 532 nm and the 532 nm HSRL Rayleigh signal are shown
in conjunction with the respective vibration-rotation Raman
signals at 387 and 607 nm, respectively, if available. We do
not compare signals at 407 nm (water-vapor Raman signals),
neither do we show water-vapor and temperature retrievals,
since these observations are currently not within the scope of
EARLINET.
Quantitative comparisons are presented for selected measurement periods from each campaign. The periods were
chosen such that the instruments showed a satisfactory performance; i.e., teething troubles as typical in the beginning
of a campaign had already been solved. Mainly nighttime
cases were considered, so that a complete evaluation of all
measurement channels and all delivered products was possible. Note that the focus of EARLINET is on the provision
of aerosol extinction and backscatter profiles independently
derived with high accuracy from Raman lidar measurements
during nighttime, since most of the Raman channels in the
network cannot be operated in the presence of strong daylight. It was ensured that the atmospheric conditions had
been stable over the measurement period, which is most
likely the case during nighttime, and allow for unambiguous comparisons. Therefore, the profiles were also checked
for the presence of a considerable amount of particles over
a large height range as well as clear-air signatures in the
far range representing Rayleigh conditions. Generally, cases
with optically thick clouds were excluded. Figures illustrating point-by-point comparisons are presented for EARLI09
only, whereas tables provide results of mean systematic deviations in selected height ranges for all intercomparison campaigns. Interested readers may find additional information on
different intercomparison campaigns in the ACTRIS QA reports; see, e.g., Deliverables D2.5 and D2.11 of the ACTRIS
FP7 project provided at http://fp7.actris.eu/language/en-GB/
Members/Deliverables.aspx.
4.1
Comparisons at signal level
Figures 3 and 4 show comparison results for the EARLI09
session of 25 May 2009, 21:00–23:00 UTC. On that day, Saharan dust layers were present up to about 6.5 km height
and provided a good opportunity for detailed comparisons
of aerosol products over a large height range. A cirrus cloud
layer occurred between 11 and 13.5 km height. The left panels of Figs. 3 and 4 show the signals at 355, 387, and 607 nm
and the total, cross-polarized, and parallel-polarized 532 nm
signals, respectively. The right panels of both figures present
the relative signal deviations from the common reference.
The applied methodology follows the explanations in Sect. 3
exactly (see Fig. 2).
The different geometrical overlap functions of the various systems and channels are clearly visible. Whereas nearrange channels based on a small telescope and a wide field
www.atmos-meas-tech.net/9/1001/2016/
1011
of view reach a complete overlap at a few hundred meters
above ground, channels based on a large-sized telescope and
a small field of view obtain full geometrical overlap between
1 and 3.5 km. The latter channels are usually well suited for
observations in the lower stratosphere up to 20–30 km height,
whereas the near-range channels are typically limited to measurements in the troposphere, as can be seen in the figures
from the large fluctuations due to low signal-to-noise ratios
above the cirrus layer. In order to account for the different
observation ranges, for each channel a valid range is defined
within which the mean deviations from the reference are calculated. The minimum valid range is the height below which
the signal has a systematic relative deviation of > 0.1 from
the reference profile, usually due to incomplete overlap. The
maximum valid range is the height above which the mean
relative deviation from the reference profile is > 0.1 over a
height interval of 2 km, usually when the detection limit is
reached. This upper boundary is determined by the instrument parameters as well as by the actual atmospheric conditions, in particular the optical depth. In the present case, the
attenuation of the signals by the cirrus cloud deck leads to
generally lower maximum valid ranges compared to observations under clear conditions.
Tables 3 and 4 show the valid range and the mean relative
signal deviation for different height ranges for the EARLI09
case of 25 May 2009 as well as for all other comparison campaigns. The height ranges are defined from the lowest valid
range to 2.5 km (R1, typically covering the planetary boundary layer), from 2.5 to 6 km (R2, representing the lower troposphere), from 6 to 12 km (R3, representing the upper troposphere), and from 12 km to the highest valid range (R4, indicating the system performance in the lower stratosphere).
If the lower valid range is above 2.5 km and/or the upper
valid range is below 12 km, the averaging is applied accordingly to the respective valid ranges, and the excluded ranges
(R1 . . . R4) are indicated as not valid (n.v.). As mentioned
above, the concept of a common reference was applied only
in EARLI09. For all other campaigns the deviations are calculated with respect to the reference system or, for stratospheric heights and when the reference system was at the
detection limit, with respect to the Rayleigh profile derived
from radiosonde observations.
Regarding EARLI09 Figs. 3 and 4 and Tables 3 and 4
show a good agreement for almost all systems. Within the
valid range the mean systematic signal deviations are, with
few exceptions, well below ±5 % and typically in the range
of ± 2 %. Best agreement is found in the lower troposphere
(R2). In this range, the mean deviations are mostly below
1 %. Largest deviations are obtained in the lowest and highest
ranges, close to the boundaries which define the valid range,
and can thus be attributed to the effects of incomplete overlap or low signal-to-noise ratio. A clear bias due to obvious
system misalignment was found for the CAML micropulse
lidar from Ispra (is01, see Fig. 4 and Table 4). Since this
commercial system is sealed, no technical corrections by the
Atmos. Meas. Tech., 9, 1001–1023, 2016
1012
U. Wandinger et al.: EARLINET intercomparison campaigns
Table 3. Valid range and mean systematic deviation of signals at 355, 387, and 607 nm in four height ranges: R1 (lowest valid range–2.5 km),
R2 (2.5–6 km), R3 (6–12 km), and R4 (12 km–highest valid range). n.v. denotes not valid. Rec. denotes receiver range.
Lidar
Rec.
Valid
range,
km
hh01
far
near
low
2.8–14.4
0.7–14.4
1.0–14.4
0.3–12.5
0.2–16.0
0.3–30.0
0.4–20.0
1.3–30.0
0.8–15.0
0.5–25.0
0.6–7.0
1.9–30.0
0.6–28.0
n.v.
−1.9
+3.0
−0.3
+2.3
−1.7
+0.8
+0.3
+1.2
−1.1
−6.1
−5.7
−1.5
−1.2
+0.0
+0.2
−0.9
+0.4
−0.0
+0.3
+0.4
+1.1
−0.4
−2.6
−0.7
−0.3
−1.7
−0.5
+0.1
+4.9
+3.4
−0.2
+0.5
−1.1
+0.4
+0.8
+9.2
+2.1
+1.1
−8.5
+2.8
+5.8
n.v.
−0.6
−0.4
+7.2
−0.1
+0.4
−1.5
n.v.
−1.9
−2.9
2.5–14.4
0.7–14.4
–
0.3–18.0
0.2–16.0
0.3–30.0
0.7–14.0
1.3–30.0
0.8–15.0
0.5–25.0
–
1.9–30.0
0.8–28.0
n.v.
−1.7
–
+2.5
+1.3
−0.4
−3.5
−3.5
+1.8
+0.0
–
−5.3
−6.4
−0.5
+0.1
–
+0.4
+0.3
+0.0
+0.2
−0.2
+0.8
+0.2
–
−0.6
−1.4
−3.8
+1.3
–
−0.8
−1.5
−0.3
−3.5
−0.9
−1.2
+1.4
–
−0.6
+3.2
−12.0
−7.5
–
+2.0
−3.6
−4.2
−4.2
−1.9
−7.4
−0.4
–
−0.3
+1.6
–
–
–
0.3–18.0
–
0.3–12.0
0.5–15.0
1.5–30.0
0.8–15.0
0.4–15.0
–
1.3–30.0
0.3–12.0
–
–
–
+0.7
–
+4.4
−2.8
−4.9
+2.1
+9.1
–
−3.5
−1.4
–
–
–
+0.2
–
+0.5
−0.2
−0.6
+0.7
+1.8
–
−0.2
−0.5
–
–
–
−0.5
–
−1.5
−0.6
+0.3
−1.6
−1.3
–
+1.2
+2.9
–
–
–
+1.1
–
n.v.
+2.1
−1.6
−5.0
−7.0
–
−1.5
n.v.
1.2–30.0
2.8–30.0
0.7–12.0
0.5–30.0
+3.3
n.v.
+1.6
+7.0
−0.2
+3.8
+2.2
+3.0
+0.1
+1.6
−3.7
+0.5
+0.3
−0.7
n.v.
−0.3
0.3–30.0
2.1–30.0
–
0.9–30.0
+2.6
+7.8
n.v.
−2.2
−0.8
+3.5
n.v.
−2.1
+0.9
−0.1
n.v.
−0.1
+0.1
−1.2
n.v.
−0.7
1.3–20.0
1.0–30.0
–
0.8–30.0
−7.3
−1.4
n.v.
−3.7
−2.2
−2.3
n.v.
−3.4
+0.3
+0.2
n.v.
−0.3
+1.6
−0.7
n.v.
−1.9
ms01
mu01
po01
mi01
le01
le02
bu01
gp01
ca01
far
near
gr01
ev01
ma01
ba02
Mean systematic deviation, %
355 nm (total)
R1
R2
R3
R4
Valid
range,
km
Mean systematic deviation, %
387 nm
R1
R2
R3
R4
Valid
range,
km
Mean systematic deviation, %
607 nm
R1
R2
R3
R4
an01
0.3–8.0
−1.4
−0.3
−4.0
n.v.
0.5–7.0
−1.7
−0.5
+0.2
n.v.
–
n.v.
n.v.
n.v.
n.v.
la01∗
0.3–13.0
+0.9
+0.2
−1.1
+2.7
0.3–13.0
+2.0
+0.0
−0.2
+2.0
–
–
–
–
–
na01
0.7–18.0
−0.4
−0.8
−1.0
−1.4
0.7–18.0
−1.7
−0.1
−1.6
−8.1
0.7–12.0
−3.0
−1.5
+8.6
n.v.
lc01
0.9–13.0
+0.2
+0.3
+0.7
−3.6
0.3–15.0
−1.4
−0.4
+0.6
−2.0
0.3–12.0
−3.4
−0.7
+5.2
n.v.
∗ XeF excimer laser; the wavelengths are 351 nm and 382 nm.
Table 4. Valid range and mean systematic deviation of signals at 532 nm (total), 532 nm (cross-polarized), and 532 nm (parallel-polarized)
in four height ranges: R1 (lowest valid range–2.5 km), R2 (2.5–6 km), R3 (6–12 km), and R4 (12 km–highest valid range). n.v. denotes not
valid. NA denotes not available. Rec. denotes receiver range.
Lidar
Rec.
Valid
range,
km
hh01
far
near
low
pol
2.0–12.5
1.6–14.4
2.0–14.4
–
0.3–13.5
0.3–20.0
0.4–20.0
1.2–28.0
0.8–15.0
(1.5–12.0)
0.4–25.0
2.5–12.0
1.4–26.0
0.2–25.0
–
−0.6
+4.7
+8.9
–
+1.5
−1.2
−2.0
+2.0
+0.5
+5.1
+4.3
n.v.
−3.9
−0.3
–
+2.6
+0.7
+1.9
–
−0.0
−0.1
−0.8
−0.3
−0.2
+3.6
−0.3
+2.6
−0.3
−0.1
–
−8.5
+0.3
−1.7
–
−0.1
−0.2
+0.4
+2.6
−0.6
−15.0
+0.1
−0.1
+0.4
+1.6
–
n.v.
+1.1
+5.6
–
+2.6
+5.9
−1.2
+5.4
−7.3
n.v.
−3.6
n.v.
−5.2
+0.8
–
–
–
–
0.5–14.4
0.3–18.0
0.3–20.0
–
1.3–30.0
–
–
0.4–30.0
–
–
–
0.5–25.0
–
–
–
+16.0
+5.2
+3.8
–
−11.0
–
–
+9.3
–
–
–
+3.8
–
–
–
+10.0
+6.2
+2.7
–
−6.5
–
–
−6.1
–
–
–
−1.8
–
–
–
−38.0
−21.0
−4.3
–
+31.0
–
–
+34.0
–
–
–
+16.0
–
–
–
−140.0
−30.0
−10.0
–
+2.0
–
–
−1.9
–
–
–
−7.5
–
–
–
0.5–14.4
0.3–14.0
0.3–20.0
0.4–20.0
–
–
–
0.4–20.0
–
–
–
0.3–25.0
–
–
–
+10.0
+1.1
−1.6
+0.3
–
–
–
+4.7
–
–
–
−5.8
–
–
–
+1.7
−0.7
−0.8
+0.7
–
–
–
+0.5
–
–
–
−1.4
–
–
–
−6.2
−0.3
+0.4
−0.7
–
–
–
−0.8
–
–
–
+1.7
–
–
–
−40.0
−3.5
+3.1
+0.3
–
–
–
−5.6
–
–
–
+1.2
gr01
ev01
ma01
ba02
0.5–30.0
1.1–30.0
0.3–25.0
2.0–30.0
−4.3
−1.2
−3.3
−8.3
−1.7
+1.1
−0.3
−3.3
+1.0
+1.5
+0.3
+1.6
−0.4
−1.9
−2.9
−0.6
1.0–30.0
–
–
–
−3.4
NA
–
–
+3.8
NA
–
–
+0.1
NA
–
–
−0.4
NA
–
–
0.5–30.0
–
–
–
−2.6
–
–
–
−0.7
–
–
–
−0.2
–
–
–
−0.3
–
–
–
an01
–
–
–
–
–
0.5–7.0
+0.9
−10.0
+3.7
n.v.
1.0–12.0
−5.7
−0.1
−8.4
n.v.
sf01∗
1.3–10.0
0.2–12.0
+0.4
+0.5
−5.0
+0.3
−7.7
+7.5
n.v.
n.v.
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
na01
1.0–15.0
−0.3
−0.2
−2.0
−3.3
1.0–13.0
+2.6
−2.2
+0.9
+9.1
−0.8
−0.1
−1.9
+2.9
lc01
1.2–15.0
+3.9
+0.4
+0.2
+5.6
–
–
–
–
–
–
–
–
–
ms01
po01
mi01
le01
le02
is01
bu01
gp01
ca01
sf02
far
near
pol
Mean systematic deviation, %
532 nm (total)
R1
R2
R3
R4
Valid
range,
km
–
–
Mean systematic deviation, %
532 nm (cross-polarized)
R1
R2
R3
R4
Valid
range,
km
–
–
1.0–15.0
–
Mean systematic deviation, %
532 nm (parallel-polarized)
R1
R2
R3
R4
∗ CuBr laser; the wavelength is 510 nm.
Atmos. Meas. Tech., 9, 1001–1023, 2016
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
1013
(b)
(a)
Range-corrected signal, a.u.
Relative deviation from common reference
0.5
10
Common reference
po01-xtg
mu01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
hh01-ltp
hh01-ntp
hh01-ftp
bu01-xtg
ca01-ntg
ca01-ftg
gp01-xta
rs09052503
1
0.1
0.01
0.005
0.4
0.3
0.2
0.1
0.0
–0.1
–0.2
–0.3
–0.4
–0.5
0
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
20
25
30
0
1
2
3
4
5
6
Height above ground, km
(c)
8
9 10 11 12 13 14 15
20
25
30
20
25
30
20
25
30
(d)
0.5
Relative deviation from common reference
Range-corrected signal, a.u.
6
Common reference
po01-xtg
mu01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
hh01-ntp
hh01-ntp (RR354)
hh01-ntp (RR353)
hh01-ftp
hh01-ftp (RR354)
hh01-ftp (RR353)
bu01-xtg
ca01-ntg
ca01-ftg
rs09052503
1
0.1
0.01
0.005
0.4
0.3
0.2
0.1
0.0
–0.1
–0.2
–0.3
–0.4
–0.5
0
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
20
25
30
0
1
2
3
4
5
6
Height above ground, km
7
8
9 10 11 12 13 14 15
Height above ground, km
(e)
(f)
5
0.5
Relative deviation from common reference
Range-corrected signal, a.u.
7
Height above ground, km
Common reference
po01-xtp
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
le01-xtp (RR530)
le01-xtp (RR529)
hh01-ntp (RR530)
hh01-ntp (RR529)
hh01-ftp (RR530)
hh01-ftp (RR529)
bu01-xtg
ca01-ntg
ca01-ftg
gp01-xta (RY532)
rs09052503
1
0.1
0.01
0.4
0.3
0.2
0.1
0.0
–0.1
–0.2
–0.3
–0.4
–0.5
0.001
0
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
Height above ground, km
20
25
30
0
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
Height above ground, km
Figure 3. Comparison of range-corrected signals at (a) 355 nm, (c) 387 nm, and (e) 607 nm and their deviations from the common reference
(b, d, f). The measurement was performed during EARLI09 on 25 May 2009, 21:00–23:00 UTC.
www.atmos-meas-tech.net/9/1001/2016/
Atmos. Meas. Tech., 9, 1001–1023, 2016
1014
U. Wandinger et al.: EARLINET intercomparison campaigns
(a)
(b)
20
10
Range-corrected signal, a.u.
Relative deviation from common reference
0.5
Common reference
po01-xsg
ms01-xsa
mi01-xta
le02-xtp
le01-xtp
is01-xtp
hh01-ltp
hh01-ntp
hh01-ftp
bu01-xsg
ca01-ntg
ca01-ftg
gp01-xta
rs09052503
1
0.1
0.01
0.4
0.3
0.2
0.1
0.0
–0.1
–0.2
–0.3
–0.4
–0.5
0.003
0
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
20
25
30
0
1
2
3
4
5
6
Height above ground, km
(c)
9 10 11 12 13 14 15
20
25
30
20
25
30
20
25
30
1.2
Relative deviation from common reference
Common reference
po01-xcg
ms01-xca
le01-xcp
hh01-fca
bu01-xcg
ca01-fcg
rs09052503
10
Range-corrected signal, a.u.
8
(d)
30
1
0.1
0.01
1.0
0.8
0.6
0.4
0.2
0.0
–0.2
–0.4
–0.6
–0.8
–1.0
–1.2
0.002
0
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
20
25
0
30
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
Height above ground, km
Height above ground, km
(e)
(f)
6
0.5
Relative deviation from common reference
Range-corrected signal, a.u.
7
Height above ground, km
Common reference
po01-xpg
ms01-xpa
mi01-xpp
hh01-fpa
bu01-xpg
ca01-fpg
rs09052503
1
0.1
0.01
0.4
0.3
0.2
0.1
0.0
–0.1
–0.2
–0.3
–0.4
–0.5
0.003
0
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
Height above ground, km
20
25
30
0
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15
Height above ground, km
Figure 4. Same as Fig. 3, but for (a, b) total, (c, d) cross-polarized, and (e, f) parallel-polarized signals at 532 nm.
Atmos. Meas. Tech., 9, 1001–1023, 2016
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
operators were possible, and the lidar could not be validated
during the campaign. The reason for the misalignment is a
temperature sensitivity of the telescope, which implies defocusing and thus different overlap functions with changing
temperature.
Other deviations seen in Figs. 3 and 4 are not considered as major quality deficits, since they are usually known
and considered in the data evaluation procedures. For instance, the rotational Raman signals (curves with symbols)
deviate because they obtain a larger attenuation than the
vibration-rotation signals (due to the shorter wavelength of
the backscattered light) and have a temperature dependence.
The spread of the 532 nm cross-polarized signals in Fig. 4c
and d is caused by the different suppression of co-polarized
radiation due to different polarizers applied in the systems. In
this case, the common reference is probably not closest to the
truth. The effects are accounted for in the polarization calibration (see Bravo-Aranda et al., 2016 and Belegante et al.,
2016). Regarding the somewhat larger deviations within the
cirrus cloud, we have to consider that inhomogeneities may
influence the signals due to the slightly different pointing of
the systems. Nevertheless, polarization-dependent transmission effects are also visible as in the case of the PollyXT system from Leipzig (le02) at 355 nm (see Fig. 3a and b). Such
effects need to be quantified and corrected for as explained
by Mattis et al. (2009) and Freudenthaler (2016).
The results provided for SPALI10 in Tables 3 and 4
are taken from two observational periods on 25 October 2010, 22:15–23:59 UTC (systems ev01, ma01, ba02),
and 4 November 2010, 20:00–20:30 UTC (gr01), because an
alignment problem of the Granada system (gr01) could be
solved only late during the campaign. Nevertheless, the more
favorable conditions during the longer measurement period
on 25 October 2010 were chosen for the comparison of the
other systems. In general, the mean systematic deviations are
somewhat larger for SPALI10 than for EARLI09. The campaign suffered from bad weather conditions and thus a limited number of suitable comparison periods. Misalignment
errors – which often occur in the beginning of the campaigns,
in particular when systems had been transported before –
could not be completely solved during SPALI10. In the case
of the PAOLI system from Évora (ev01; see Tables 3 and
4) the reason for the large deviations in the height ranges
R1 and R2, which are due to a very large range of incomplete overlap, could be identified only when the system was
back at Évora. It was found that the field stop was not exactly positioned on the receiver optical axis, possibly because
of damage during transport. In addition, it was not possible
to obtain successful intercomparisons for all channels during SPALI10. In particular, the signals of the CIEMAT lidar
from Madrid (ma01) showed electronic disturbances, varying from day to day, which prevented the Raman channels at
387 and 607 nm from being verified.
For comparing the Alomar Tropospheric Lidar (an01) with
the reference system (hh01) during ALI09, several meawww.atmos-meas-tech.net/9/1001/2016/
1015
surement periods on 4 November 2009, between 09:00 and
16:30 UTC, have been investigated. The optical receiver of
the Alomar system had been changed considerably before the
campaign. It turned out that the setup was not stable throughout the ALI09 campaign. Readjustments were necessary for
each session, and it was not possible to obtain a good performance of all channels at the same time. In addition, signal
offsets both in analog and photon-counting channels were
observed temporarily, which hint to external electronic disturbances in the laboratory environment. From Tables 3 and
4 it can be seen that reasonable agreement with the reference
system could be achieved for the 355, 387, and 532 nm channels up to the mid-troposphere during selected time periods.
The 607 nm signal could not be validated. Consequences are
discussed in Sect. 5.
The systems at Sofia were compared during different sessions of SOLI10. Because of the different pointing angles of
the lidars, the reference system (mu01) was operated under
a zenith angle of 0◦ on 10 October 2010, 17:34–17:56 UTC,
to compare the sf01 system, and under a zenith angle of 60◦
on 10 October 2010, 20:06–20:46 UTC, to compare the sf02
system. Good comparison results were achieved. A minor
height shift in the signal of the sf02-system was attributed
to a wrong trigger-delay correction. Deviations in the upper
troposphere for both systems are due to low signal-to-noise
ratios caused by the low system power in the case of sf01 and
the large zenith angle in the case of sf02.
The intercomparison period selected for the LALI12 campaign is 15 September 2012, 22:31–23:39 UTC. Excellent
performance of the L’Aquila UV lidar (la01) was obtained.
Mean systematic deviations from the signals of the reference
system (mu01) were ≤ 2 % throughout the entire observational range (see Table 3).
For NALI13, the selected measurement period to calculate the numbers presented in Tables 3 and 4 is 17–18 October 2013, 23:03–00:06 UTC. The MALIA system (na01)
compared very well to the reference system (po01), and no
significant deviations in any of the channels were obtained.
The Lecce system (lc01) was compared to the reference
system (po01) during LELI13 on 22 October 2013, 19:01–
20:01 UTC. In principle, also here the results were satisfactory. Some smaller biases were identified in the near range
and attributed to geometrical effects due to focussing of the
signals onto the photocathodes. Further discussion of system
deficiencies found in all campaigns and proposed solutions
is provided in Sect. 5.
4.2
Comparisons at product level
Figures 5 and 6 show comparison results for particle extinction and backscatter coefficients at 355 and 532 nm,
respectively, for the EARLI09 session of 25 May 2009,
21:00–23:00 UTC, derived from the signals presented in
Figs. 3 and 4. Particle extinction coefficients were calculated after the Raman method (Ansmann et al., 1990) from
Atmos. Meas. Tech., 9, 1001–1023, 2016
1016
U. Wandinger et al.: EARLINET intercomparison campaigns
(a)
(b)
0.25
0.10
0.15
0.10
Absolute deviation from reference, km
Extinction coefficient, km
-1
0.20
-1
Common reference
po01-xtg
mu01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
hh01-ntp
hh01-ntp (RR)
hh01-ftp
hh01-ftp (RR)
bu01-xtg
ca01-ntg
ca01-ftg
0.05
0.00
–0.05
0.05
0.00
–0.05
–0.10
0
1
2
3
4
5
6
7
8
9
10
0
1
2
3
Height above ground, km
(c)
0.003
0.002
7
8
9
10
8
9
10
8
9
10
-1
-1
Absolute deviation from reference, km sr
-1
0.004
-1
Backscatter coefficient, km sr
6
0.0010
Common reference
po01-xtg
mu01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
hh01-ntp
hh01-ntp (RR)
hh01-ftp
hh01-ftp (RR)
bu01-xtg
ca01-ntg
ca01-ftg
0.001
0.000
–0.001
0.0008
0.0006
0.0004
0.0002
0.0000
–0.0002
–0.0004
–0.0006
–0.0008
–0.0010
0
1
2
3
4
5
6
7
8
9
10
0
1
2
Height above ground, km
3
4
5
6
7
Height above ground, km
(e)
(f)
0.004
-1
0.0010
0.003
0.002
-1
Absolute deviation from reference, km sr
Common reference
po01-xtg
mu01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
hh01-ntp
hh01-ftp
hh01-ltp
bu01-xtg
ca01-ntg
ca01-ftg
gp01-xta
-1
-1
5
(d)
0.005
Backscatter coefficient, km sr
4
Height above ground, km
0.001
0.000
–0.001
0.0008
0.0006
0.0004
0.0002
0.0000
–0.0002
–0.0004
–0.0006
–0.0008
–0.0010
0
1
2
3
4
5
6
7
Height above ground, km
8
9
10
0
1
2
3
4
5
6
7
Height above ground, km
Figure 5. Comparison of particle extinction coefficients (a) and particle backscatter coefficients derived using the Raman (c) and Fernald
methods (e), respectively, at 355 nm and their absolute deviations from the common reference (b, d, e). The measurement was performed
during EARLI09 on 25 May 2009, 21:00–23:00 UTC.
Atmos. Meas. Tech., 9, 1001–1023, 2016
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
1017
Table 5. Valid range and mean systematic deviation of particle extinction coefficients at 355 and 532 nm in four height ranges: R1 (lowest
valid range–2.5 km), R2 (2.5–6 km), R3 (6–12 km), and R4 (12 km–highest valid range). n.v. denotes not valid. Rec. denotes receiver range.
Lidar
Rec.
Valid
range,
km
hh01
far
near
far
3.6–11.3
1.5–11.3
0.4–16.5
0.4–14.5
0.8–17.5
1.2–10.0
1.9–20.0
1.7–15.0
1.1–22.0
–
2.9–22.0
1.2–22.0
n.v.
−7.1
+9.5
+0.6
+1.1
−13.0
−13.0
−1.9
+4.7
–
n.v.
−10.0
+2.6
−1.1
+3.6
+3.6
+1.8
+3.5
−0.3
+6.1
+2.4
–
−2.5
−9.6
+5.9
+1.4
−0.6
+3.0
−0.4
+0.3
+0.4
+0.8
−2.1
–
−1.8
−2.0
n.v.
n.v.
+4.7
−5.1
+3.5
n.v.
+0.6
+7.8
+5.8
–
+2.5
+3.9
–
1.2–11.3
0.8–16.5
–
0.8–11.3
1.0–11.3
3.0–22.5
1.7–15.0
1.0–15.0
2.5–10.8
2.6–22.0
0.6–11.3
n.v.
−6.5
−7.6
–
+1.8
−17.0
n.v.
−1.4
+14.0
n.v.
n.v.
−8.3
n.v.
+0.2
+0.8
–
+6.7
−4.3
−4.8
+3.9
+12.0
+2.0
−2.8
−4.5
n.v.
+2.9
+0.1
–
+8.3
+8.5
+0.9
+2.1
−2.6
−0.3
−1.5
−3.7
n.v.
n.v.
+6.4
–
n.v.
n.v.
+2.3
+5.8
−1.6
n.v.
+0.8
n.v.
gr01
ev01
ma01
ba02
1.0–14.0
2.3–15.0
–
1.0–13.0
+8.1
+9.3
n.v.
+3.2
+1.1
+6.6
n.v.
+5.7
+16.0
+2.4
n.v.
+0.3
+24.0
+1.9
n.v.
+8.9
1.5–10.0
1.5–8.0
–
1.2–6.0
+0.3
+5.9
n.v.
−14.0
−6.2
−3.0
n.v.
−19.0
−8.2
+24.0
n.v.
n.v.
n.v.
n.v.
n.v.
n.v.
la01∗
0.5–13.0
+7.5
+2.1
+2.1
+5.6
–
–
–
–
–
na01
1.6–14.0
−4.3
+1.9
+0.8
+6.7
1.5–8.0
−6.4
+9.3
−0.1
n.v.
lc01
1.0–8.0
−3.8
−1.6
−1.2
n.v.
0.9–6.0
−11.0
−1.9
n.v.
n.v.
ms01
mu01
po01
mi01
le01
le02
bu01
gp01
ca01
far
near
Mean systematic deviation,
10−3 km−1 , 355 nm extinction
coefficient
R1
R2
R3
R4
Valid
range,
km
Mean systematic deviation,
10−3 km−1 , 532 nm extinction
coefficient
R1
R2
R3
R4
∗ XeF excimer laser; the wavelength is 351 nm.
vibration-rotation Raman signals at 387 and 607 nm, respectively, or from HSRL Rayleigh and rotational Raman signals if available. In the latter case, the two temperaturedependent signals were added in order to get a profile that
is nearly temperature-independent. Particle backscatter coefficients were calculated using both the Raman (Ansmann
et al., 1992) and the Fernald methods (Fernald, 1984). The
reference height range is 7–10 km for the Raman solutions
and 9.4–10 km for the Fernald solutions. The profiles were
calibrated in these height ranges to mean values of 4 × 10−5
and 2.5 × 10−5 km−1 sr−1 at 355 and 532 nm, respectively.
A lidar ratio of 55 sr was chosen in the Fernald algorithm. A
gliding average with increasing window length over height
was applied in both extinction and backscatter retrievals. For
extinction, the window length increases from 180 m below
1 km height to 2.7 km in the stratosphere. For backscatter,
the resolution is 60 m up to 3.4 km and increases to 2.7 km
in the stratosphere. All retrievals and comparisons were done
without any correction of the individual overlap function.
Tables 5 and 6 show the valid range and the mean absolute deviations from the reference of particle extinction
and backscatter coefficients, respectively, for the predefined height ranges for all comparison campaigns. For the
backscatter coefficients, the Raman solutions are considered
www.atmos-meas-tech.net/9/1001/2016/
whenever possible. Otherwise, the Fernald solutions are used
(italic numbers in the table). Profiles are considered to be
valid when they systematically deviate from the reference by
< 0.01 km−1 at the low end and by < 0.025 km−1 at the far
end of the profile in the case of extinction. For the backscatter coefficients, the limit is set to 3 × 10−4 km−1 sr−1 at both
ends. These values are of the order of the statistical measurement errors (see fluctuation of the deviations in the right
panels of Figs. 5 and 6) and typically about 10–20 % of the
particle extinction and backscatter coefficients measured in
distinct aerosol layers (see left panels of Figs. 5 and 6). If
cirrus clouds were present, e.g., in the case of the EARLI09
example, these height ranges were excluded from the averages in Tables 5 and 6 because of the heterogeneity of the
products due to different measurement geometry. Because of
different pointing and various fields of view, not only different volumes are detected, but also the influence of specular
reflection and multiple scattering on the products varies from
instrument to instrument.
Extinction retrievals (see Figs. 5a, b and 6a, b) clearly
show the influence of the different overlap functions. The
curves are cut at the lower valid range defined for Raman signals from which they were derived (see Table 3). The lower
valid range for the particle extinction coefficient is clearly
Atmos. Meas. Tech., 9, 1001–1023, 2016
1018
U. Wandinger et al.: EARLINET intercomparison campaigns
(a)
(b)
0.10
0.10
0.05
-1
Common reference
po01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
le01-xtp (RR)
hh01-ntp (RR)
bu01-xtg
ca01-ntg
ca01-ftg
gp01-xta (RY)
Absolute deviation from reference, km
Extinction coefficient, km
-1
0.15
0.00
–0.05
0.05
0.00
–0.05
–0.10
0
1
2
3
4
5
6
7
8
9
10
0
1
2
Height above ground, km
6
7
8
9
10
8
9
10
8
9
10
0.0010
0.002
0.001
-1
-1
Absolute deviation from reference, km sr
Common reference
po01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
le01-xtp (RR)
hh01-ftp (RR)
hh01-ntp (RR)
bu01-xtg
ca01-ntg
ca01-ftg
gp01-xta
-1
-1
Backscatter coefficient, km sr
5
(d)
0.003
0.000
0.0008
0.0006
0.0004
0.0002
0.0000
–0.0002
–0.0004
–0.0006
–0.0008
–0.0010
–0.001
0
1
2
3
4
5
6
7
8
9
0
10
1
2
3
4
5
6
7
Height above ground, km
Height above ground, km
(e)
(f)
0.0010
-1
0.003
0.002
0.001
-1
Absolute deviation from reference, km sr
Common reference
po01-xtg
ms01-xtg
mi01-xtp
le02-xtp
le01-xtp
is01-xtp
hh01-ntp
hh01-ftp
hh01-ltp
bu01-xtg
ca01-ntg
ca01-ftg
gp01-xta
-1
-1
4
Height above ground, km
(c)
Backscatter coefficient, km sr
3
0.000
0.0008
0.0006
0.0004
0.0002
0.0000
–0.0002
–0.0004
–0.0006
–0.0008
–0.0010
–0.001
0
1
2
3
4
5
6
7
Height above ground, km
8
9
10
0
1
2
3
4
5
6
7
Height above ground, km
Figure 6. Same as Fig. 5, but for 532 nm.
Atmos. Meas. Tech., 9, 1001–1023, 2016
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
1019
Table 6. Valid range and mean systematic deviation of backscatter coefficients at 355 and 532 nm in four height ranges: R1 (lowest valid
range–2.5 km), R2 (2.5–6 km), R3 (6–12 km), and R4 (12 km–highest valid range). Italicized numbers indicate Fernald retrievals; all other
numbers belong to the Raman method. n.v. denotes not valid. Rec. denotes receiver range.
Lidar
Rec.
Valid
range,
km
hh01
far
near
far
3.5–13.4
0.7–13.4
0.3–11.3
0.3–19.0
0.3–30.0
0.8–10.0
1.3–28.0
0.8–17.0
–
0.5–30.0
0.6–2.5
1.0–30.0
0.8–28.0
n.v.
+0.1
−85.0
−3.0
−21.0
+40.0
+18.0
−0.5
–
−22.0
−9.0
−14.0
+64.0
−12.0
+7.6
−42.0
−3.3
−1.8
+3.7
+2.3
+2.4
–
+1.6
n.v.
−4.2
+21.0
+0.1
+1.7
−0.1
+0.7
+0.2
+10.0
−0.1
+1.5
–
−0.1
n.v.
+0.6
+1.4
+4.6
+1.4
n.v.
+5.0
+6.6
n.v.
+5.3
+9.7
–
+2.5
n.v.
−0.1
+0.1
–
–
0.3–17.0
–
0.3–30.0
0.5–11.0
1.3–30.0
0.8–15.0
(1.5–11.5)
0.5–15.0
2.5–11.3
1.4–27.0
0.4–30.0
gr01
ev01
ma01
ba02
0.7–30.0
0.2–25.0
1.0–8.0
0.5–28.0
−2.6
+13.0
−0.5
+0.1
−0.3
+2.3
+2.1
+5.2
−6.3
+0.0
−13.0
−0.6
−4.0
+6.9
n.v.
+3.2
sf01a
sf02
–
–
–
–
–
–
–
–
–
–
la01b
0.2–13.0
+8.5
+0.2
−6.3
−0.3
–
na01
0.9–18.0
+12.0
+2.8
+0.7
+6.7
lc01
0.8–13.0
+13.0
+5.2
+2.0
+8.2
ms01
mu01
po01
mi01
le01
le02
is01
bu01
gp01
ca01
far
near
Mean systematic deviation,
10−5 km−1 sr−1 , 355 nm backscatter
coefficient
R1
R2
R3
R4
Valid
range,
km
Mean systematic deviation,
10−5 km−1 sr−1 , 532 nm backscatter
coefficient
R1
R2
R3
R4
–
–
+9.3
–
−11.0
−4.2
+15.0
−9.4
+ 34.0
−12.0
n.v.
+3.8
+13.0
–
–
+1.3
–
−1.6
−6.5
−6.4
−6.4
+33.0
−6.7
−5.5
+1.5
+3.6
–
–
+0.7
–
+3.4
+0.6
−0.8
−1.2
+7.3
+1.1
−0.4
+0.4
+0.7
–
–
−12.0
–
−14.0
n.v.
+0.6
−1.5
n.v.
+4.9
n.v.
−0.6
−5.2
0.4–15.0
0.2–14.0
0.3–25.0
0.2–14.0
+4.7
−2.2
−3.0
−2.2
+0.0
+0.9
−0.7
+0.9
−1.8
−4.0
−0.2
−2.7
−5.7
+17.0
−0.3
+19.0
1.5–10.0
0.5–12.0
+14.0
−0.7
+3.6
+5.6
+1.0
+2.2
n.v.
n.v.
–
–
–
–
0.7–12.0
+6.6
+1.5
−2.3
n.v.
1.0–12.0
+12.0
+0.1
−1.3
n.v.
a CuBr laser; the wavelength is 510 nm,
b XeF excimer laser; the wavelength is 351 nm.
higher and above 0.8 km for most systems, even when the receiver is optimized for the near range. When complete overlap is reached, the mean systematic deviations of the particle extinction coefficients are small and typically well below ±0.01 km−1 throughout the troposphere. Signal noise
is the dominating source of uncertainty then, in particular
at 532 nm above 3–4 km height, where several curves show
large fluctuations (see Fig. 6a, e.g., mi01, bu01).
Backscatter coefficients can be derived closer towards
the ground than extinction coefficients. In the Raman retrieval the overlap effect cancels out when both the elasticbackscatter signal and the Raman signal have the same geometrical overlap function. However, since differences in optical imaging and signal nonlinearities may occur in the near
range, this compensation does not work in all cases, as can
be seen from Figs. 5c, d and 6c, d. Another reason for the
spread of the curves towards the ground is the identical calibration in a common reference range. When the signals are
disturbed in this range, the calibration fails and the whole
profile is corrupted. For instance, the particle backscatter cowww.atmos-meas-tech.net/9/1001/2016/
efficient at 355 nm of the MULIS system (ms01) is shifted
towards values that are too low because of an analog signal
distortion (positive offset) in the calibration range (7–10 km).
The Fernald solutions for the particle backscatter coefficients at 355 and 532 nm are shown in Figs. 5e, f and 6e, f,
respectively. The curves are cut at the lower end according to the valid range of the elastic-backscatter signals provided in Tables 3 and 4, i.e., when the relative signal deviation from the reference becomes > 10 %. It can be seen that
close to this lower boundary, the particle backscatter coefficients are derived with an acceptable absolute deviation of
< 7 × 10−4 and < 3 × 10−4 km−1 sr−1 at 355 and 532 nm,
respectively. Analog signal distortions in the reference range
lead to offsets here as well (see, e.g., gp01 in Fig. 5e, f). In
Fig. 6e, f the result of the retrieval for the CAML lidar from
Ispra (is01) is also shown. The misalignment discussed above
leads to a bias of up to 5 × 10−4 km−1 sr−1 in the free troposphere. Beside those few exceptions, for which the reasons
could be identified, the mean systematic deviation of the particle backscatter coefficients, from Raman as well as Fernald
Atmos. Meas. Tech., 9, 1001–1023, 2016
1020
retrievals, is < 2 × 10−4 km−1 sr−1 above the defined minimum valid range (see Table 6).
5
Discussion
The EARLINET intercomparisons performed between 2009
and 2013 provided a broad insight into the level of quality
that has been reached in the network after nearly 15 years of
operation. The decision to perform comparisons not only at
product level as done before (Matthias et al., 2004) but also at
signal level, based on a common preprocessing, allowed for
a much deeper analysis of individual measurement channels
and potential system failures. For the majority of the detection channels, mean systematic deviations from the reference
over predefined height ranges were below ±1–2 % in the upper PBL and the free troposphere. Particle backscatter and
extinction coefficients could then be retrieved with an accuracy of better than ±2 × 10−4 km−1 sr−1 and ±0.01 km−1 ,
respectively. These values are well below the quality margins of ±5 × 10−4 km−1 sr−1 and ±0.05 km−1 defined by
Matthias et al. (2004) and of the order of 10 % of typical particle backscatter and extinction values observed in the PBL.
Some of the signals showed higher systematic biases,
which were further investigated. Typical reasons were misalignment errors. Such errors were often observed in the beginning of the campaigns, in particular when the systems
had been transported before. This finding depicts a shortcoming of centralized intercomparison campaigns for which
many systems have to be moved and operated outside of their
normal environments. Personnel that are used to working
with stationary systems under well-defined conditions usually need some time to gain experience with the challenges of
a field campaign. Such problems do not occur when on-site
comparisons with a well-characterized traveling standard are
performed instead. Some of the systematic biases obtained
could be attributed to signal distortions in analog detection
channels or problems with the gluing of analog and photoncounting signals due to incorrect setting of acquisition parameters. These facts also call for expert site visits and individual training in order to check systems in their laboratory
environment and advise personnel, in particular new network
members, in specific quality-checking procedures.
Further shortcomings observed during the intercomparison campaigns, which could not be solved on-site, led to consequences regarding system upgrades or replacement. After the failure of CAML (is01) in the EARLI09 campaign,
this system was removed from the network and the station
at Ispra was upgraded with a new 3 + 2 lidar system manufactured by Raymetrics (even if the performance of CAML
could be improved after EARLI09 by operating the lidar at
a controlled temperature of 35 ± 4 ◦ C). The new lidar has
been in operation at Ispra since March 2013. Several systems with low performance in the PBL have been upgraded
with near-range receivers, among them the MARTHA sysAtmos. Meas. Tech., 9, 1001–1023, 2016
U. Wandinger et al.: EARLINET intercomparison campaigns
tem from Leipzig (le01). Also, the newest generation of
PollyXT lidars (here represented by le02) is equipped with
additional channels that allow measurements down to about
50 m height (Engelmann et al., 2015). In the CIEMAT lidar system (ma01), which suffered from mechanical instability and electronic disturbances in the two Raman channels
during SPALI10, the respective PMTs (model Hamamatsu
R928) were replaced. The new data acquisition is based on
a Licel/Hamamatsu PMT R7400P-20 for the 607 nm channel
and a Licel/Hamamatsu PMT R9880 U-110 for the 387 nm
channel, and combined analog and photon-counting detection is applied. Moreover, mechanical modifications for a
better robustness of the system were implemented. After the
LELI13 campaign in Lecce, during which some biases in the
near range of the UNILE system (lc01) had been detected,
the receiver of the multiwavelength lidar at this station was
modified and the single focussing lens in front of each detector was replaced with a collimator in order to avoid geometrical effects due to inhomogeneities of the detector surfaces.
Regarding the Alomar Tropospheric Lidar (an01), which
showed major deficiencies during ALI09, a number of measures, implemented after discussion with EARLINET experts, resulted in distinct improvements of the system. The
electrical noise, induced by the laser, could be reduced by
using a fiber coupling to achieve a galvanic separation of
the data acquisition electronics and the light source. The
cause for the poor quality of the 607 nm channel was identified as the combination of an interference filter that is too
broad and a photomultiplier with poor quantum efficiency at
this wavelength. During a system refurbishment, this channel has been removed, the main mirror was recoated, and
the crosstalk of the depolarization channels was minimized
with additional polarizing sheet filters. In addition, an automated polarization calibration unit was installed (Freudenthaler et al., 2009).
In general, dedicated intercomparison campaigns as discussed in this paper require large efforts and can thus only
be performed sporadically. Nevertheless, because of the lack
of external calibration standards for aerosol lidar observations, any instrument intercomparison is of great value for
quality assurance. Therefore, following the principle of best
scientific practice, every opportunity of cross-checking the
quality of measurements by a direct comparison of results
from collocated observations should be used. Within EARLINET and in collaboration between EARLINET and other
research projects, direct instrument intercomparisons are performed whenever possible. Regular intercomparisons take
place at sites where more than one system is available, e.g.,
because the groups own one of the reference systems in addition to their stationary lidar (Potenza, Minsk, Hamburg, Munich/Maisach) or apply other lidars in experiments outside
of EARLINET (Leipzig, Naples). Other opportunities are related to dedicated field campaigns in which several lidars often participate. In this context, comparisons with downlook-
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
ing airborne lidars may also be used to check the system performance in the near range.
6
Conclusions and outlook
In this paper, we have presented results of the EARLINET
instrument intercomparison campaigns between 2009 and
2013. During this period, about two-thirds of the EARLINET systems performed comparison measurements with
one or more reference systems. In two dedicated campaigns, EARLI09 and SPALI10, 15 instruments underwent
this quality-assurance procedure. EARLI09 also served to
qualify the reference systems that are used as traveling standards within the network. With these reference instruments,
six other systems were checked during direct station visits. Altogether, more than 100 individual measurement channels were examined, based on a common strategy of signal
preprocessing and evaluation following the principles of the
EARLINET Single Calculus Chain. In most cases, a very
good agreement of signals as well as derived aerosol products with the defined reference could be obtained. The intercomparisons have reinforced confidence in the EARLINET
data quality and allowed us to draw conclusions on necessary
system improvements for some instruments and to identify
major challenges that need to be tackled in the future.
EARLINET is a living network that is continuously in
development, both regarding the instrument level and the
network distribution. Most of the stations regularly upgrade
their systems by adding new measurement capabilities based
on recent experience, technological developments, and available funding. Thus, a complete assessment of all systems at
any time in any specific setup through intercomparison with a
reference system is not possible. Therefore, complementary
quality-assurance concepts need to be applied. EARLINET
requires regular internal system check-ups in addition to the
sporadic intercomparisons. Specific internal check-up tools,
such as the telecover test for the near range and the Rayleigh
fit for the far range, have been developed and must be applied at least once per year as well as after each major system
upgrade. These activities are discussed in detail by Freudenthaler et al. (2016).
Within the ACTRIS-2 project, which started in May 2015,
the quality-assurance efforts of EARLINET will be further improved. In order to provide a long-term, sustainable infrastructure that can serve the lidar community even
beyond EARLINET, the Lidar Calibration Centre, LiCal,
will be established. LiCal is a common effort of the EARLINET groups at INOE (Bucharest, Romania), CNR-IMAA
(Potenza, Italy), and LMU (Munich, Germany). It will provide lidar calibration services from the characterization and
optimization of single components to the assessment of complete systems through intercomparison with reference systems. The POLIS (mu01) and MUSA (po01) reference lidars will be further upgraded and used exclusively for this
www.atmos-meas-tech.net/9/1001/2016/
1021
purpose in the future. LiCal will also host a lidar training
laboratory in order to educate station personnel in applying lidar check-up tools and maintaining lidar calibration. In
this way, the long-term quality-assurance strategy of EARLINET, which has been successfully established over the past
15 years, will be further consolidated.
Acknowledgements. The financial support for EARLINET–ASOS
by the European Commission in the Sixth Framework Programme
under grant RICA-025991, for ACTRIS in the Seventh Framework Programme under grant agreement no. 262254, and for
ACTRIS-2 in HORIZON 2020 under grant agreement no. 654109,
as well as the ESA financial support under the ESRIN contract
no. 22202/09/I-EC, are gratefully acknowledged. We thank Julia
Fruntke, Christian Herold, and the technical staff of TROPOS for
the logistical support, radiosonde launches, and weather forecast
during EARLI09, as well as the Atmosphere Group of INTAMadrid for their support with the radiosounding during SPALI10.
The Universitat Politècnica de Catalunya wishes to acknowledge
the technical and logistic support of Joaquim Giner and Rubén
Tardío. The work of the INOE team was supported by grant
no. 229907 FP7-REGPOT-2008-1. The Universitat Politècnica
de Catalunya group received support from the Spanish Ministry
of Economy and Competitiveness (project TEC2012-34575) and
of Science and Innovation (project UNPC10-4E-442), as well as
from the Department of Economy and Knowledge of the Catalonia
autonomous government (grant 2014 SGR 583).
Edited by: R. Ferrare
References
Althausen, D., Engelmann, R., Baars, H., Heese, B., Ansmann,
A., Müller, D., and Komppula, M.: Portable Raman Lidar
PollyXT for automated profiling of aerosol backscatter, extinction, and depolarization, J. Atmos. Ocean. Tech., 26, 2366–2378,
doi:10.1175/2009JTECHA1304.1, 2009.
Ansmann, A., Riebesell, M., and Weitkamp, C.: Measurement of
atmospheric aerosol extinction profiles with a Raman lidar, Opt.
Lett., 15, 746–748, doi:10.1364/OL.15.000746, 1990.
Ansmann, A., Riebesell, M., Wandinger, U., Weitkamp, C., Voss,
E., Lahmann, W., and Michaelis, W.: Combined Raman elasticbackscatter lidar for vertical profiling of moisture, aerosol extinction, backscatter, and lidar ratio, Appl. Phys. B, 55, 18–28,
doi:10.1007/BF00348608, 1992.
Apituley, A., Wilson, K. M., Potma, C., Volten, H., and de Graaf,
M.: Performance Assessment and Application of Caeli – A highperformance Raman lidar for diurnal profiling of Water Vapour,
Aerosols and Clouds, in: Proceedings of the 8th International
Symposium on Tropospheric Profiling, Delft, the Netherlands,
19–23 October 2009, edited by: Apituley, A., Russchenberg, H.
W. J., and Monna, W. A. A., ISBN 978-90-6960-233-2, 2009.
Barnaba, F., Putaud, J. P., Gruening, C., dell’Acqua, A., and Dos
Santos, S.: Annual cycle in co-located in situ, total-column, and
height-resolved aerosol observations in the Po Valley (Italy): Implications for ground-level particulate matter mass concentration
Atmos. Meas. Tech., 9, 1001–1023, 2016
1022
estimation from remote sensing, J. Geophys. Res., 115, D19209,
doi:10.1029/2009JD013002, 2010.
Belegante, L., Nicolae, D., Nemuc, A., Talianu, C. and Derognat,
C.: Retrieval of the boundary layer height from active and passive
remote sensors. Comparison with a NWP model, Acta Geophys.,
62, 276–289, doi:10.2478/s11600-013-0167-4, 2014.
Belegante, L., Bravo-Aranda, J. A., Freudenthaler, V., Nicolae, D.,
Nemuc, A., Alados-Arboledas, L., Amodeo, A., Pappalardo, G.,
D’Amico, G., Engelmann, R., Baars, H., Wandinger, U., Papayannis, A., Kokkalis, P., and Pereira, S. N.: Experimental assessment of the lidar polarizing sensitivity, Atmos. Meas. Tech.
Discuss., doi:10.5194/amt-2015-337, in review, 2016.
Böckmann, C., Wandinger, U., Ansmann, A., Bösenberg, J.,
Amiridis, V., Boselli, A., Delaval, A., De Tomasi, F., Frioud, M.,
Hågård, A., Horvat, M., Iarlori, M., Komguem, L., Kreipl, S.,
Larchevêque, G., Matthias, V., Papayannis, A., Pappalardo, G.,
Rocadembosch, F., Rodriguez, J. A., Schneider, J., Shcherbakov,
V., and Wiegner, M.: Aerosol lidar intercomparison in the framework of the EARLINET project. 2. Aerosol backscatter algorithms, Appl. Opt., 43, 977–989, doi:10.1364/AO.43.000977,
2004.
Böckmann, C., Mironova, I., Müller, D., Scheidenbach, L.,
and Nessler, R.: Microphysical aerosol parameters from
multiwavelength lidar, J. Opt. Soc. Am. A, 22, 518–528,
doi:10.1364/JOSAA.22.000518, 2005.
Bravo-Aranda, J. A., Belegante, L., Freudenthaler, V., AladosArboledas, A., Nicolae, D., Granados-Muñoz, M. J., GuerreroRascado, J. L., Amodeo, A., D’Amico, G., Engelmann, R., Pappalardo, G., Kokkalis, P., Mamouri, R., Papayannis, A., NavasGuzmán, F., Olmo, F. J., Wandinger, U., and Haeffelin, M.:
Assessment of lidar depolarization uncertainty by means of
a polarimetric lidar simulator, Atmos. Meas. Tech. Discuss.,
doi:10.5194/amt-2015-339, in review, 2016.
D’Amico, G., Amodeo, A., Baars, H., Binietoglou, I., Freudenthaler, V., Mattis, I., Wandinger, U., and Pappalardo, G.: EARLINET Single Calculus Chain – overview on methodology and
strategy, Atmos. Meas. Tech., 8, 4891–4916, doi:10.5194/amt-84891-2015, 2015.
D’Amico, G., Amodeo, A., Mattis, I., Freudenthaler, V., and Pappalardo, G.: EARLINET Single Calculus Chain – technical – Part
1: Pre-processing of raw lidar data, Atmos. Meas. Tech., 9, 491–
507, doi:10.5194/amt-9-491-2016, 2016.
Engelmann, R., Kanitz, T., Baars, H., Heese, B., Althausen, D.,
Skupin, A., Wandinger, U., Komppula, M., Stachlewska, I. S.,
Amiridis, V., Marinou, E., Mattis, I., Linné, H., and Ansmann,
A.: EARLINET Raman Lidar PollyXT: the neXT generation, Atmos. Meas. Tech. Discuss., 8, 7737–7780, doi:10.5194/amtd-87737-2015, 2015.
Engelmann, R., Guerrero-Rascado, J. L., Alados-Arboledas, L.,
Wandinger, U., Freudenthaler, V., Baars, H., Mattis, I., Groß,
S., Pappalardo, G., Amodeo, A., D’Amico, G., Giunta, A.,
Chaikovsky, A., Osipenko, F., Slesar, A., Nicolae, D., Belegante,
L., Serikov, I., Linné, H., Jansen, F., Apituley, A., Wilson, K.,
Trickl, T., and Rocadenbosch, F.: Calibrated backscatter measurements at 1064 nm with lidar: Techniques used in EARLINET
and ACTRIS, Atmos. Meas. Tech. Discuss., in preparation, 2016.
Fernald, F. G.: Analysis of atmospheric lidar observations: some comments, Appl. Opt., 23, 652–653,
doi:10.1364/AO.23.000652, 1984.
Atmos. Meas. Tech., 9, 1001–1023, 2016
U. Wandinger et al.: EARLINET intercomparison campaigns
Freudenthaler, V.: About the effects of polarising optics on lidar
signals and the 190-calibration, Atmos. Meas. Tech. Discuss.,
doi:10.5194/amt-2015-338, in review, 2016.
Freudenthaler, V., Esselborn, M., Wiegner, M., Heese, B., Tesche,
M., Ansmann, A., Müller, D., Althausen, D., Wirth, M., Fix, A.,
Ehret, G., Knippertz, P., Toledano, C., Gasteiger, J., Garhammer,
M., and Seefeldner, M.: Depolarization ratio profiling at several wavelengths in pure Saharan dust during SAMUM 2006,
Tellus B, 61, 165–179, doi:10.1111/j.1600-0889.2008.00396.x,
2009.
Freudenthaler, V., Linné, H., Chaikovsky, A., Groß, S., and Rabus,
D.: Internal quality assurance tools, Atmos. Meas. Tech. Discuss., in preparation, 2016.
Freudenthaler, V., Seefeldner, M., Groß, S., and Wandinger, U.: Accuracy of linear depolarisation ratios in clean air ranges measured
with POLIS-6 at 355 and 532 nm, 27th International Laser Radar
Conference, 6–10 July 2015, New York City, USA, available at:
http://epub.ub.uni-muenchen.de/24942/ (last access: 18 February
2016), 2015.
Guerrero-Rascado, J. L., Ruiz, B., and Alados-Arboledas, L.: Multispectral Lidar characterization of the vertical structure of Saharan dust aerosol over Southern Spain, Atmos. Environ., 42,
2668–2681, doi:10.1016/j.atmosenv.2007.12.062, 2008.
Guerrero-Rascado, J. L., Olmo, F. J., Avilés-Rodríguez, I., NavasGuzmán, F., Pérez-Ramírez, D., Lyamani, H., and Alados Arboledas, L.: Extreme Saharan dust event over the southern
Iberian Peninsula in september 2007: active and passive remote
sensing from surface and satellite, Atmos. Chem. Phys., 9, 8453–
8469, doi:10.5194/acp-9-8453-2009, 2009.
Kumar, D., Rocadenbosch, F., Sicard, M., Comerón, A., Muñoz, C.,
Lange, D., Tomás, S., and Gregorio, E.: Six-channel polychromator design and implementation for the UPC elastic/Raman lidar,
in: Proc. of SPIE, Vol. 8182, edited by: Singh, U. N. and Pappalardo, G., pp. 81820W–1–81820W–10, International Society
for Optics and Photonics, doi:10.1117/12.896305, 2011.
Leblanc, T., Walsh, T. D., McDermid, I. S., Toon, G. C., Blavier,
J.-F., Haines, B., Read, W. G., Herman, B., Fetzer, E., Sander, S.,
Pongetti, T., Whiteman, D. N., McGee, T. G., Twigg, L., Sumnicht, G., Venable, D., Calhoun, M., Dirisu, A., Hurst, D., Jordan,
A., Hall, E., Miloshevich, L., Vömel, H., Straub, C., Kampfer,
N., Nedoluha, G. E., Gomez, R. M., Holub, K., Gutman, S.,
Braun, J., Vanhove, T., Stiller, G., and Hauchecorne, A.: Measurements of Humidity in the Atmosphere and Validation Experiments (MOHAVE)-2009: overview of campaign operations and
results, Atmos. Meas. Tech., 4, 2579–2605, doi:10.5194/amt-42579-2011, 2011.
Madonna, F., Amodeo, A., Boselli, A., Cornacchia, C., Cuomo,
V., D’Amico, G., Giunta, A., Mona, L., and Pappalardo, G.:
CIAO: the CNR-IMAA advanced observatory for atmospheric
research, Atmos. Meas. Tech., 4, 1191–1208, doi:10.5194/amt4-1191-2011, 2011.
Matthias, V., Freudenthaler, V., Amodeo, A., Balin, I., Balis, D.,
Bösenberg, J., Chaikovsky, A., Chourdakis, G., Comeron, A.,
Delaval, A., De Tomasi, F., Eixmann, R., Hågård, A., Komguem,
L., Kreipl, S., Matthey, R., Rizi, V., Rodrigues, J., Wandinger,
U., and Wang, X.: Aerosol lidar intercomparison in the framework of the EARLINET project. 1. Instruments, Appl. Opt., 43,
961–976, doi:10.1364/AO.43.000961, 2004.
www.atmos-meas-tech.net/9/1001/2016/
U. Wandinger et al.: EARLINET intercomparison campaigns
Mattis, I., Ansmann, A., Müller, D., Wandinger, U., and Althausen,
D.: Multiyear aerosol observations with dual-wavelength Raman
lidar in the framework of EARLINET, J. Geophys. Res., 109,
D13203, doi:10.1029/2004JD004600, 2004.
Mattis, I., Ansmann, A., Müller, D., Wandinger, U., and Althausen, D.: Systematic error of lidar profiles caused by
a polarization-dependent receiver transmission: Quantification
and error correction scheme, Appl. Opt., 48, 2742–2751,
doi:10.1364/AO.48.002742, 2009.
Mattis, I., D’Amico, G., Baars, H., Amodeo, A., Madonna, F., and
Ialori, M.: EARLINET Single Calculus Chain – technical – Part
2: Calculations of optical products, Atmos. Meas. Tech. Discuss.,
submitted, 2016.
Müller, D., Wandinger, U., and Ansmann, A.: Microphysical particle parameters from extinction and backscatter lidar data by inversion with regularization: Theory, Appl. Opt., 38, 2346–2357,
doi:10.1364/AO.38.002346, 1999.
Müller, D., Böckmann, C., Kolgotin, A., Schneidenbach, L., Chemyakin, E., Rosemann, J., Znak, P., and Romanov, A.: Microphysical particle properties derived from inversion algorithms
developed in the framework of EARLINET, Atmos. Meas.
Tech. Discuss., 8, 12823–12885, doi:10.5194/amtd-8-128232015, 2015.
Nair, P. J., Godin-Beekmann, S., Froidevaux, L., Flynn, L. E., Zawodny, J. M., Russell III, J. M., Pazmiño, A., Ancellet, G.,
Steinbrecht, W., Claude, H., Leblanc, T., McDermid, S., van Gijsel, J. A. E., Johnson, B., Thomas, A., Hubert, D., Lambert,
J.-C., Nakane, H., and Swart, D. P. J.: Relative drifts and stability of satellite and ground-based stratospheric ozone profiles
at NDACC lidar stations, Atmos. Meas. Tech., 5, 1301–1318,
doi:10.5194/amt-5-1301-2012, 2012.
Nemuc, A., Vasilescu, J., Talianu, C., Belegante, L., and Nicolae,
D.: Assessment of aerosol’s mass concentrations from measured
linear particle depolarization ratio (vertically resolved) and simulations, Atmos. Meas. Tech., 6, 3243–3255, doi:10.5194/amt-63243-2013, 2013.
www.atmos-meas-tech.net/9/1001/2016/
1023
Pappalardo, G., Amodeo, A., Pandolfi, M., Wandinger, U., Ansmann, A., Bösenberg, J., Matthias, V., Amiridis, V., De Tomasi,
F., Frioud, M., Iarlori, M., Komguem, L., Papayannis, A., Rocadenbosch, F., and Wang, X.: Aerosol lidar intercomparison in the
framework of the EARLINET project. 3. Raman lidar algorithm
for aerosol extinction, backscatter and lidar ratio, Appl. Opt., 43,
5370–5385, doi:10.1364/AO.43.005370, 2004.
Pappalardo, G., Amodeo, A., Apituley, A., Comeron, A., Freudenthaler, V., Linné, H., Ansmann, A., Bösenberg, J., D’Amico,
G., Mattis, I., Mona, L., Wandinger, U., Amiridis, V., AladosArboledas, L., Nicolae, D., and Wiegner, M.: EARLINET: towards an advanced sustainable European aerosol lidar network,
Atmos. Meas. Tech., 7, 2389–2409, doi:10.5194/amt-7-23892014, 2014.
Perrone, M. R., De Tomasi, F., and Gobbi, G. P.: Vertically resolved
aerosol properties by multi-wavelength lidar measurements, Atmos. Chem. Phys., 14, 1185–1204, doi:10.5194/acp-14-11852014, 2014.
Rizi, V., Iarlori, M., Rocci, G., and Visconti, G.: Raman lidar observations of cloud liquid water, Appl. Opt., 43, 6440–6453,
doi:10.1364/AO.43.006440, 2004.
Schmidt, J., Wandinger, U., and Malinka, A.: Dual-fieldof-view Raman lidar measurements for the retrieval of
cloud microphysical properties, Appl. Opt., 52, 2235–2247,
doi:10.1364/AO.52.002235, 2013.
Stoyanov, D., Grigorov, I., Kolarov, G., Peshev, Z., and Dreischuh,
T.: LIDAR Atmospheric Sensing by Metal Vapor and Nd:YAG
Lasers, in Advanced Photonic Sciences, edited by: Fadhali, M.,
ISBN 978-953-307-927-1, InTech, 2011.
Veselovskii, I., Kolgotin, A., Griaznov, V., Müller, D., Wandinger,
U., and Whiteman, D. N.: Inversion with regularization for
the retrieval of tropospheric aerosol parameters from multiwavelength lidar sounding, Appl. Opt., 41, 3685–3699,
doi:10.1364/AO.41.003685, 2002.
Atmos. Meas. Tech., 9, 1001–1023, 2016
Fly UP